COMPARISON OF RIGHT VENTRICULAR IMPEDANCE, PULSE PRESSURE AND MAXIMAL DP/DT FOR DETERMINATION OF HEMODYNAMIC STABILITY OF VENTRICULAR ARRHYTHMIAS ASSOCIATED WITH CORONARY-ARTERY DISEASE

被引:10
作者
WOOD, MA
ELLENBOGEN, KA
KAPADIA, K
LU, B
VALENTA, H
机构
[1] HUNTER HOLMES MCGUIRE VET ADM MED CTR,DEPT MED,DIV CARDIOL,111J,1201 BROAD ROCK BLVD,RICHMOND,VA 23249
[2] VIRGINIA COMMONWEALTH UNIV,MED COLL VIRGINIA,RICHMOND,VA 23298
[3] TELECTR PACING SYST,DENVER,CO
关键词
D O I
10.1016/0002-9149(90)90484-I
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
By monitoring hemodynamic parameters, a future generation of automatic implantable defibrillators will provide tiered therapy of ventricular arrhythmias according to the associated hemodynamic compromise. Changes in intracardiac impedance permit beat-to-beat assessment of ventricular volumes and make this parameter attractive as a rapid discriminator of hemodynamic compromise during arrhythmias. Beat-to-beat changes in right ventricular (RV) impedance were measured before, during and after 27 episodes of ventricular tachyarrhythmias induced in 17 men (64 ± 7 years, mean ± standard deviation; left ventricular ejection fraction 41 ± 11%). Impedance was measured using a tripolar lead system and was compared to ventricular tachycardia cycle length, RV pulse pressure and maximum systolic RV dP dt as indicators of systemic hemodynamic compromise. The average decreases in systolic blood pressure and mean arterial pressure during ventricular tachycardia were 48 ± 23% and 46 ± 26%, respectively (mean ± standard deviation; p < 0.001 for each). Right ventricular impedance decreased an average 39 ± 22% from its baseline value (p < 0.001) during ventricular tachycardia. The percent change in impedance from baseline during ventricular tachycardia correlated significantly with the percent decrease in systolic and mean arterial pressure (r = 0.45 and 0.42, respectively; both p < 0.05). Right ventricular dP dt correlated the most poorly of all parameters with changes in blood pressure while impedance X RV pulse pressure correlated best with changes in mean and systolic pressure (r ≥ 0.82, p < 0.001). © 1990.
引用
收藏
页码:575 / 582
页数:8
相关论文
共 11 条
[1]   CONTINUOUS MEASUREMENT OF LEFT-VENTRICULAR VOLUME IN ANIMALS AND HUMANS BY CONDUCTANCE CATHETER [J].
BAAN, J ;
VANDERVELDE, ET ;
DEBRUIN, HG ;
SMEENK, GJ ;
KOOPS, J ;
VANDIJK, AD ;
TEMMERMAN, D ;
SENDEN, J ;
BUIS, B .
CIRCULATION, 1984, 70 (05) :812-823
[2]   CONTINUOUS STROKE VOLUME AND CARDIAC-OUTPUT FROM INTRAVENTRICULAR DIMENSIONS OBTAINED WITH IMPEDANCE CATHETER [J].
BAAN, J ;
JONG, TTA ;
KERKHOF, PLM ;
MOENE, RJ ;
VANDIJK, AD ;
VANDERVELDE, ET ;
KOOPS, J .
CARDIOVASCULAR RESEARCH, 1981, 15 (06) :328-334
[3]   ACCURACY OF VOLUME MEASUREMENT BY CONDUCTANCE CATHETER IN ISOLATED, EJECTING CANINE HEARTS [J].
BURKHOFF, D ;
VANDERVELDE, E ;
KASS, D ;
BAAN, J ;
MAUGHAN, WL ;
SAGAWA, K .
CIRCULATION, 1985, 72 (02) :440-447
[4]   CONTINUOUS RIGHT VENTRICULAR VOLUME ASSESSMENT BY CATHETER MEASUREMENT OF IMPEDANCE FOR ANTITACHYCARDIA SYSTEM CONTROL [J].
KHOURY, D ;
MCALISTER, H ;
WILKOFF, B ;
SIMMONS, T ;
RUDY, Y ;
MCCOWAN, R ;
MORANT, V ;
CASTLE, L ;
MALONEY, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (12) :1918-1926
[5]   INSTANTANEOUS MEASUREMENT OF LEFT AND RIGHT VENTRICULAR STROKE VOLUME AND PRESSURE-VOLUME RELATIONSHIPS WITH AN IMPEDANCE CATHETER [J].
MCKAY, RG ;
SPEARS, JR ;
AROESTY, JM ;
BAIM, DS ;
ROYAL, HD ;
HELLER, GV ;
LINCOLN, W ;
SALO, RW ;
BRAUNWALD, E ;
GROSSMAN, W .
CIRCULATION, 1984, 69 (04) :703-710
[6]   ELECTRODE CONFIGURATIONS FOR A RESPIRATORY IMPEDANCE MEASUREMENT SUITABLE FOR RATE RESPONSIVE PACING [J].
NAPPHOLTZ, T ;
VALENTA, H ;
MALONEY, J ;
SIMMONS, T .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (06) :960-964
[7]   CONTINUOUS VENTRICULAR VOLUME ASSESSMENT FOR DIAGNOSIS AND PACEMAKER CONTROL [J].
SALO, RW ;
PEDERSON, BD ;
OLIVE, AL ;
LINCOLN, WC ;
WALLNER, TG .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (06) :1267-1272
[8]   OBSERVATIONS ON BLOOD-FLOW RELATED ELECTRICAL-IMPEDANCE CHANGES IN RIGID TUBES [J].
VISSER, KR ;
LAMBERTS, R ;
KORSTEN, HHM ;
ZIJLSTRA, WG .
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1976, 366 (2-3) :289-291
[9]   PREDICTORS OF INDUCIBLE SUSTAINED VENTRICULAR TACHYARRHYTHMIAS IN PATIENTS WITH CORONARY-ARTERY DISEASE [J].
VORPERIAN, VR ;
GITTELSOHN, AM ;
VELTRI, EP .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 13 (03) :637-645
[10]   RIGHT VENTRICULAR VOLUMETRY BY CATHETER MEASUREMENT OF CONDUCTANCE [J].
WOODARD, JC ;
BERTRAM, CD ;
GOW, BS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (04) :862-870