[O-15]H2O, [O-15]O-2 AND [O-15]CO GAS-PRODUCTION, MONITORING AND QUALITY-CONTROL SYSTEM

被引:10
作者
JACKSON, JR [1 ]
DEMBOWSKI, BS [1 ]
EHRENKAUFER, RL [1 ]
MCINTYRE, E [1 ]
REIVICH, M [1 ]
机构
[1] UNIV PENN,CEREBROVASC RES CTR,PHILADELPHIA,PA 19104
关键词
D O I
10.1016/0969-8043(93)90181-9
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The O-15 gas production system selects the gas production mode, monitors gas delivery conditions and performs quality control needed during the production and delivery of continuous flow radioactive gas for equilibrium and bolus PET studies. Target gas switches choose the appropriate valve positioning for [O-15]H2O, [O-15]O2 or [O-15]CO production while flow rate, pressure and radioactivity are constantly monitored. The system also includes an IBM PC interfaced ps chromatograph which provides chemical and radiochemical quality control by on-line, real time, gas sampling of the delivery line.
引用
收藏
页码:631 / 634
页数:4
相关论文
共 17 条
[1]  
BERRIDGE MS, 1990, APPL RADIAT ISOTOPES, V41, P1173
[2]  
Clark J. C., 1975, SHORT LIVED RADIOACT
[3]  
CLARK JC, 1987, APPL CLIN ISOT, V38, P1211
[4]   QUANTITATIVE MEASUREMENT OF REGIONAL CEREBRAL BLOOD-FLOW AND OXYGEN-METABOLISM IN MAN USING O-15 AND POSITRON EMISSION TOMOGRAPHY - THEORY, PROCEDURE, AND NORMAL VALUES [J].
FRACKOWIAK, RSJ ;
LENZI, GL ;
JONES, T ;
HEATHER, JD .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1980, 4 (06) :727-736
[5]   QUANTITATIVE MEASUREMENT OF LOCAL CEREBRAL BLOOD-FLOW IN HUMANS BY POSITRON COMPUTED-TOMOGRAPHY AND O-15-WATER [J].
HUANG, SC ;
CARSON, RE ;
HOFFMAN, EJ ;
CARSON, J ;
MACDONALD, N ;
BARRIO, JR ;
PHELPS, ME .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1983, 3 (02) :141-153
[6]  
IWATA R, 1988, APPL RADIAT ISOTOPES, V39, P1207
[7]   TANDEM VANDEGRAAFF ACCELERATOR PRODUCTION OF POSITRON LABELED RADIOPHARMACEUTICALS FOR ROUTINE CLINICAL USE [J].
JONES, SC ;
ROBINSON, GD ;
MCINTYRE, E .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1984, 35 (08) :721-729
[8]   CEREBRAL BLOOD-FLOW WITH THE CONTINUOUS INFUSION OF OXYGEN-15-LABELED WATER [J].
JONES, SC ;
GREENBERG, JH ;
DANN, R ;
ROBINSON, GD ;
KUSHNER, M ;
ALAVI, A ;
REIVICH, M .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1985, 5 (04) :566-575
[9]   CONTINUOUS INHALATION OF OXYGEN-15 FOR ASSESSING REGIONAL OXYGEN EXTRACTION IN BRAIN OF MAN [J].
JONES, T ;
CHESLER, DA ;
TERPOGOSSIAN, MM .
BRITISH JOURNAL OF RADIOLOGY, 1976, 49 (580) :339-343
[10]   MEASUREMENT OF CEREBRAL BLOOD-FLOW USING BOLUS INHALATION OF (CO2)-O-15 AND POSITRON EMISSION TOMOGRAPHY - DESCRIPTION OF THE METHOD AND ITS COMPARISON WITH THE (CO2)-O-15 CONTINUOUS INHALATION METHOD [J].
KANNO, I ;
LAMMERTSMA, AA ;
HEATHER, JD ;
GIBBS, JM ;
RHODES, CG ;
CLARK, JC ;
JONES, T .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1984, 4 (02) :224-234