THE ECAT EXACT HR - PERFORMANCE OF A NEW HIGH-RESOLUTION POSITRON SCANNER

被引:483
作者
WIENHARD, K
DAHLBOM, M
ERIKSSON, L
MICHEL, C
BRUCKBAUER, T
PIETRZYK, U
HEISS, WD
机构
[1] MAX PLANCK INST NEUROL FORSCH, SCH MED, D-50931 COLOGNE, GERMANY
[2] UNIV CALIF LOS ANGELES, LOS ANGELES, CA USA
[3] KAROLINSKA HOSP, PET LAB, STOCKHOLM, SWEDEN
[4] UCL, LOUVAIN, BELGIUM
关键词
EMISSION COMPUTED TOMOGRAPHY; APPARATUS AND EQUIPMENT; PERFORMANCE EVALUATION; POSITRON EMISSION TOMOGRAPHY; TECHNIQUES;
D O I
10.1097/00004728-199401000-00023
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The ECAT EXACT HR is a newly designed CTI-Siemens PET scanner with high spatial resolution. Its physical performance with respect to resolution, count rate efficiency, and scatter was investigated and evaluated with phantom studies. Materials and Methods: The new tomograph consists of three rings of 112 BGO block detectors (50 mm x 23 mm x 30 mm deep) each, covering an axial field of view of 15 cm with a patient port of 56 cm diameter. Each block is sawed into an 8 x 7 matrix giving 24 detector rings with 784 crystals each. Results: Total sensitivity for a 20 cm cylinder phantom is 177 kcps/mu Ci/ml in two-dimensional (2D) mode and increases to 1.46 Mcps/mu Ci/ml in 3D mode. Count rate performance was investigated for different low energy discriminator thresholds. Smaller detector blocks improve noise equivalent counts by similar to 50% compared with the EXACT system both in 2D and in 3D mode. Scatter fractions vary in 2D from 0.09 to 0.13 for energy thresholds from 450 to 250 keV for line sources in a 20 cm diameter phantom. In 3D mode an increase of scatter by a factor of 3 is observed. Transaxial spatial resolution varies from 3.6 mm full width at half-maximum (FWHM) at the center to 4.5 mm FWHM tangentially and 7.4 mm FWHM radially at R = 20 cm. Average axial resolution changes from 4.0 mm FWHM at center to 6.7 mm FWHM at R = 20 cm. Conclusion: Due to its special properties, the EXACT HR can be equally applied to routine clinical brain and whole-body imaging and to noninvasive experimental studies of regional tracer concentrations in medium-sized animals
引用
收藏
页码:110 / 118
页数:9
相关论文
共 13 条
  • [1] CORRECTION FOR SCATTERED RADIATION IN A RING DETECTOR POSITRON CAMERA BY INTEGRAL TRANSFORMATION OF THE PROJECTIONS
    BERGSTROM, M
    ERIKSSON, L
    BOHM, C
    BLOMQVIST, G
    LITTON, J
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1983, 7 (01) : 42 - 50
  • [2] A MULTICRYSTAL 2-DIMENSIONAL BGO DETECTOR SYSTEM FOR POSITRON EMISSION TOMOGRAPHY
    CASEY, ME
    NUTT, R
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) : 460 - 463
  • [3] CUTLER PD, 1992, J NUCL MED, V33, P595
  • [4] OPTIMIZATION OF PET INSTRUMENTATION FOR BRAIN ACTIVATION STUDIES
    DAHLBOM, M
    CHERRY, SR
    ERIKSSON, L
    HOFFMAN, EJ
    WIENHARD, K
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (04) : 1048 - 1054
  • [5] A POSITRON TOMOGRAPH WITH 600-BGO CRYSTALS AND 2.6 MM RESOLUTION
    DERENZO, SE
    HUESMAN, RH
    CAHOON, JL
    GEYER, AB
    MOSES, WW
    UBER, DC
    VULETICH, T
    BUDINGER, TF
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) : 659 - 664
  • [6] DERENZO SE, 1993, ANN NUCL MED, V7, pS3
  • [7] 3-D PHANTOM TO SIMULATE CEREBRAL BLOOD-FLOW AND METABOLIC IMAGES FOR PET
    HOFFMAN, EJ
    CUTLER, PD
    DIGBY, WM
    MAZZIOTTA, JC
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) : 616 - 620
  • [8] KARP JS, 1991, J NUCL MED, V32, P2342
  • [9] NOISE IDENTIFICATION AND REMOVAL IN POSITRON IMAGING-SYSTEMS
    KING, PH
    HUBNER, K
    GIBBS, W
    HOLLOWAY, E
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1981, 28 (01) : 148 - 151
  • [10] MOSES WW, 1993, J NUCL MED, V34, pP101