TIMING PROPERTIES OF PHILIPS XP2020UR-PHOTOMULTIPLIER

被引:17
作者
MOSZYNSKI, M
机构
[1] Centre de Recherches Nucléaires, IN2P3-CNRS / Université Louis Pasteur, F-67037 Strasbourg Cedex
关键词
D O I
10.1016/0168-9002(93)90988-T
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The timing properties of scintillation counters consisting of small NE111 plastic scintillators coupled to the XP2020UR photomultipliers have been studied in comparison to that based on the XP2020. The time jitter of both the photomultipliers, their photoelectron yield for the NE111 plastics and the time resolution with Co-60 and Na-22 sources were measured. A high time resolution was achieved with the XP2020UR equal to 90 ps for Co-60 (at 1 MeV) and 140 ps for Na-22 (at 290 keV) energy settings, about 30% and 20% better respectively than those observed with the XP2020. The large improvement of the time resolution is due to: 30% lower time jitter, about 25% higher photoelectron yield and for the Co-60 energy settings also to a reduced influence of the space charge effect in the XP2020UR. The normalised time resolution to the photoelectron number, measured with Na-22 source, is better by a factor of 1.14 which corresponds well to the improvement of the time jitter in the new XP2020UR photomultiplier.
引用
收藏
页码:269 / 275
页数:7
相关论文
共 24 条
[1]  
ABRAMS GS, 1909, NUCL INSTRUM METH A, V281, P55
[2]   TIMING IMPROVED BY THE USE OF DYNODE SIGNALS STUDIED WITH DIFFERENT SCINTILLATORS AND PHOTOMULTIPLIERS [J].
BENGTSON, B ;
MOSZYNSKI, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 204 (01) :129-140
[3]  
BENGTSON B, UNPUB
[4]  
BERTOLACCINI M, 1968, P NUCL ELECTR S VERS
[5]   THE MARK-III TIME-OF-FLIGHT SYSTEM [J].
BROWN, JS ;
BURNETT, TH ;
COOK, V ;
DELPAPA, C ;
DUNCAN, AL ;
MOCKETT, PM ;
SLEEMAN, JC ;
WILLUTZKI, HJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 221 (03) :503-522
[6]   FAST TIMING WITH HAMAMATSU R2083Q PHOTOMULTIPLIERS [J].
DEVRIES, J ;
ZECCA, A ;
BRUSA, RS ;
GRISENTI, RG ;
OSS, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 275 (01) :194-196
[7]   TESTS OF CONVENTIONAL AND MICROCHANNEL PLATE PHOTOMULTIPLIERS AS COMPONENTS OF A SCINTILLATION TIME-OF-FLIGHT SYSTEM [J].
GILES, RT ;
PIPKIN, FM ;
WOLINSKI, JP .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 252 (01) :41-52
[8]   TIMING PROPERTIES OF LONG SCINTILLATION-COUNTERS BASED ON SCINTILLATING FIBERS [J].
KUHLEN, M ;
MOSZYNSKI, M ;
STROYNOWSKI, R ;
WICKLUND, E ;
MILLIKEN, B .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 301 (02) :223-229
[9]   COMPARISON OF THE TIMING PROPERTIES OF THE NEW PHILIPS COMPONENTS (AMPEREX) XP2020/UR-PHOTOMULTIPLIER AND THE XP2020-PHOTOMULTIPLIER [J].
KUHLEN, M ;
MILLIKEN, B ;
STROYNOWSKI, R ;
WICKLUND, E .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1991, 38 (05) :1052-1055
[10]  
KUHLEN M, 1989, P INT IND S SSC NEW