ADAPTATION OF PULSE OXIMETRY FOR FETAL MONITORING DURING LABOR

被引:68
作者
GARDOSI, JO
SCHRAM, CM
SYMONDS, EM
机构
[1] Perinatal Research and Monitoring Unit, Department of Obstetrics and Gynaecology, Queen's Medical Centre, Nottingham
关键词
D O I
10.1016/0140-6736(91)92928-U
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Application of pulse oximetry to intrapartum monitoring was investigated in 105 women. No adequate reading could be obtained in 44 cases. Two major sources of artifact, related to probe apposition and signal processing, were identified and excluded. The average arterial oxygen saturation from the fetal scalp was 82% (SD 6%), which is higher than has been inferred from pO2 levels. Readings below 60% or the development of an unstable baseline suggest the presence of substantial hypoxia and acidosis.
引用
收藏
页码:1265 / 1267
页数:3
相关论文
共 12 条
  • [1] FETAL HEMOGLOBIN CONCENTRATION IN NEWBORN - INDEX OF MATURITY AND AS SUPPORTIVE EVIDENCE FOR MATERNAL FETAL TRANSFUSION
    ANDREWS, BF
    WILLET, GP
    [J]. AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1965, 91 (01) : 85 - &
  • [2] DAMIANOU D, 1991, IEE ABSTR, V89, P3
  • [3] GARDOSI J, 1989, LANCET, V2, P692
  • [4] ABSORPTION CHARACTERISTICS OF HUMAN-FETAL HEMOGLOBIN AT WAVELENGTHS USED IN PULSE OXIMETRY
    HARRIS, AP
    SENDAK, MJ
    DONHAM, RT
    THOMAS, M
    DUNCAN, D
    [J]. JOURNAL OF CLINICAL MONITORING, 1988, 4 (03): : 175 - 177
  • [5] FETAL MONITORING WITH PULSE OXIMETRY
    JOHNSON, N
    JOHNSON, VA
    FISHER, J
    JOBBINGS, B
    BANNISTER, J
    LILFORD, RJ
    [J]. BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 1991, 98 (01): : 36 - 41
  • [6] ADHERENCE IN CHILDHOOD DIABETES - RESULTS OF A CONFIRMATORY FACTOR-ANALYSIS
    JOHNSON, SB
    TOMER, A
    CUNNINGHAM, WR
    HENRETTA, JC
    [J]. HEALTH PSYCHOLOGY, 1990, 9 (04) : 493 - 501
  • [7] PEAT S, 1988, LANCET, V2, P213
  • [8] ROSEN M, 1990, 1ST INT S INTR SURV
  • [9] THE EFFECT OF CAPUT SUCCEDANEUM ON OXYGEN-SATURATION MEASUREMENTS
    SCHRAM, CMH
    GARDOSI, J
    [J]. BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 1991, 98 (01): : 113 - 114
  • [10] PULSE OXIMETER AND TRANS-CUTANEOUS ARTERIAL OXYGEN MEASUREMENTS IN NEONATAL AND PEDIATRIC INTENSIVE-CARE
    SOUTHALL, DP
    BIGNALL, S
    STEBBENS, VA
    ALEXANDER, JR
    RIVERS, RPA
    LISSAUER, T