DOSIMETRY AND PHYSICAL TREATMENT PLANNING FOR IODINE EYE PLAQUE THERAPY

被引:8
作者
ALBERTI, W
POTHMANN, B
TABOR, P
MUSKALLA, K
HERMANN, KP
HARDER, D
机构
[1] UNIV ESSEN GESAMTHSCH KLINIKUM, STRAHLEN KLIN, W-4300 ESSEN 1, GERMANY
[2] UNIV GOTTINGEN, INST MED PHYS & BIOPHYS, W-3400 GOTTINGEN, GERMANY
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1991年 / 20卷 / 05期
关键词
EYE PLAQUES; I-125; SEEDS; DOSIMETRY; EYE PHANTOM;
D O I
10.1016/0360-3016(91)90209-M
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The dosimetry of eye plaques loaded with iodine-125 seeds (type 6702) was performed by means of computer calculations and measurements with thermoluminescent dosimeters (TLD). Measurements of the depth dose distribution (2-25.5 mm) along the transverse axis of a single seed were performed in water equivalent phantom material. The transverse axis attenuation and geometry factor F(r) was obtained by applying a least squares fit to the measured data. Based on the resulting radial dose function, a computer program was developed which calculates dose distributions within the eye for arbitrary loading and placement of the eye plaque. The computational results were verified by TLD measurements in an eye phantom.
引用
收藏
页码:1087 / 1092
页数:6
相关论文
共 20 条
[1]  
ANDERSON LL, 1975, HDB INTERSTITIAL BRA, P87
[2]   TWO-DIMENSIONAL DOSE DISTRIBUTION AROUND A COMMERCIAL I125 SEED [J].
BURNS, GS ;
RAESIDE, DE .
MEDICAL PHYSICS, 1988, 15 (01) :56-60
[3]   MONTE-CARLO SIMULATION OF THE DOSE DISTRIBUTION AROUND I-125 SEEDS [J].
BURNS, GS ;
RAESIDE, DE .
MEDICAL PHYSICS, 1987, 14 (03) :420-424
[4]   DOSE-RATE DETERMINATION FOR I-125 SEEDS [J].
CHIUTSAO, ST ;
ANDERSON, LL ;
OBRIEN, K ;
SANNA, R .
MEDICAL PHYSICS, 1990, 17 (05) :815-825
[5]   MONTE-CARLO DOSIMETRY FOR I-125 AND CO-60 IN EYE PLAQUE THERAPY [J].
CHIUTSAO, ST ;
OBRIEN, K ;
SANNA, R ;
TSAO, HS ;
VIALOTTI, C ;
CHANG, YS ;
ROTMAN, M ;
PACKER, S .
MEDICAL PHYSICS, 1986, 13 (05) :678-682
[7]   AN I-125 PLAQUE FOR RADIOTHERAPY OF THE EYE - MANUFACTURE AND DOSIMETRIC CONSIDERATIONS [J].
HARNETT, AN ;
THOMSON, ES .
BRITISH JOURNAL OF RADIOLOGY, 1988, 61 (729) :835-838
[8]   THE 3-DIMENSIONAL DOSE DISTRIBUTION OF I-125 SEEDS IN TISSUE [J].
HARTMANN, GH ;
SCHLEGEL, W ;
SCHARFENBERG, H .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (06) :693-699
[9]   POLYETHYLENE-BASED WATER-EQUIVALENT PHANTOM MATERIAL FOR X-RAY DOSIMETRY AT TUBE VOLTAGES FROM 10 TO 100 KV [J].
HERMANN, KP ;
GEWORSKI, L ;
MUTH, M ;
HARDER, D .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (11) :1195-1200
[10]  
HOLT JG, 1975, EC0013 MEM HOSP REP, P37