摘要
In recent years there has been a growing awareness of the possible hazards caused by anaesthetic gases in operating theatres. The laryngeal mask airway provides an alternative both to tracheal intubation and the face mask although the implications for operating theatre contamination have not been quantified. This paper describes the incidence and magnitude of exposure of theatre personnel to waste anaesthetic gases during laryngeal mask airway anaesthesia. The leakage of anaesthetic gases to the anaesthetist's breathing zone was monitored using a Bruel & Kjaer Multi Gas Monitor, Type 1302 during 50 general anaesthetics employing either spontaneous (n = 24) or controlled (n = 26) ventilation. All patients were anaesthetised with propofol, alfentanil and nitrous oxide. There was no statistically significant association between the amount of anaesthetic gas leakage and ventilation method. The laryngeal mask airway meets occupational safety requirements on nitrous oxide concentrations in the operating theatre environment. Copyright © 1992, Wiley Blackwell. All rights reserved
引用
收藏
页码:697 / 700
页数:4
相关论文
共 12 条
[1]   MONITORING OF NITROUS-OXIDE IN OPERATING-ROOMS - IDENTIFICATION OF SOURCES AND ESTIMATION OF OCCUPATIONAL EXPOSURE [J].
BORM, PJA ;
KANT, I ;
HOUBEN, G ;
VANRIJSSENMOLL, M ;
HENDERSON, PT .
JOURNAL OF OCCUPATIONAL AND ENVIRONMENTAL MEDICINE, 1990, 32 (11) :1112-1116
[2]   THE LARYNGEAL MASK - A NEW CONCEPT IN AIRWAY MANAGEMENT [J].
BRAIN, AIJ .
BRITISH JOURNAL OF ANAESTHESIA, 1983, 55 (08) :801-805
[3]  
BRAIN AIJ, 1985, ANAESTHESIA, V40, P356
[4]   FURTHER DEVELOPMENTS OF THE LARYNGEAL MASK [J].
BRAIN, AIJ .
ANAESTHESIA, 1989, 44 (06) :530-530
[5]   THE LARYNGEAL MASK AIRWAY - A STUDY OF 100 PATIENTS DURING SPONTANEOUS BREATHING [J].
BRODRICK, PM ;
WEBSTER, NR ;
NUNN, JF .
ANAESTHESIA, 1989, 44 (03) :238-241
[7]   THE LARYNGEAL MASK AIRWAY - CLINICAL APPRAISAL IN 250 PATIENTS [J].
MALTBY, JR ;
LOKEN, RG ;
WATSON, NC .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1990, 37 (05) :509-513
[8]   THE USE OF THE FIBEROPTIC LARYNGOSCOPE TO CONFIRM THE POSITION OF THE LARYNGEAL MASK [J].
PAYNE, J .
ANAESTHESIA, 1989, 44 (10) :865-865
[9]   LARYNGEAL MASK AND ANESTHETIC WASTE-GAS CONCENTRATIONS [J].
SARMA, VJ ;
LEMAN, J .
ANAESTHESIA, 1990, 45 (09) :791-792
[10]   HAFNIA A CIRCUIT - VALVELESS NON-POLLUTING ANESTHETIC SYSTEM [J].
THOMSEN, A ;
JORGENSEN, S .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1976, 20 (04) :395-404