AMPEROMETRIC DETERMINATION OF GLUCOSE IN UNDILUTED FOOD SAMPLES

被引:38
作者
AMINE, A [1 ]
PATRIARCHE, GJ [1 ]
MARRAZZA, G [1 ]
MASCINI, M [1 ]
机构
[1] DIPARTIMENTO SANITA PUBBL EPIDEMIOL & CHIM ANAL AMBIENTALE,SEZIONE CHIM ANAL,I-50121 FLORENCE,ITALY
关键词
AMPEROMETRY; FLOW SYSTEM; GLUCOSE; FOODS; SOFT DRINKS; ENZYME ELECTRODES;
D O I
10.1016/0003-2670(91)87051-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Glucose oxidase is immobilized onto a cellulose acetate membrane by glutaraldehyde linkage, and the membrane is used to cover the platinum electrode of a hydrogen peroxide sensor. A silanized polycarbonate membrane then covers the enzyme layer, and extends the linear calibration range to higher concentrations. The sensor, when incorporated into a flow-injection system, allows the determination of glucose at levels up to 1 M in soft drinks at a rate of 60 samples h-1 without sample dilution.
引用
收藏
页码:91 / 98
页数:8
相关论文
共 6 条
  • [1] KARUBE I, 1989, MRS INT M ADV MATERI, V14
  • [2] GLUCOSE ELECTROCHEMICAL PROBE WITH EXTENDED LINEARITY FOR WHOLE-BLOOD
    MASCINI, M
    SELLERI, S
    RIVIELLO, M
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1989, 7 (12) : 1507 - 1512
  • [3] MASCINI M, 1989, ANAL LETT, V22, P1429
  • [4] MASCINI M, 1989, SELECT ELECTR REV, V11, P191
  • [5] GLUCOSE ENZYME ELECTRODE WITH EXTENDED LINEARITY - APPLICATION TO UNDILUTED BLOOD MEASUREMENTS
    MULLEN, WH
    KEEDY, FH
    CHURCHOUSE, SJ
    VADGAMA, PM
    [J]. ANALYTICA CHIMICA ACTA, 1986, 183 : 59 - 66
  • [6] Turner A., 1987, BIOSENSORS FUNDAMENT