MINIMIZATION OF INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY USING ONLINE PRECONCENTRATION

被引:93
作者
HEITHMAR, EM
HINNERS, TA
ROWAN, JT
RIVIELLO, JM
机构
[1] LOCKHEED ENGN & SCI CO,LAS VEGAS,NV 89119
[2] DIONEX CHEM CORP,SUNNYVALE,CA 94088
关键词
D O I
10.1021/ac00207a018
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A semiautomated system is used to preconcentrate Ti, V, Mn, Fe, Co, Ni, Cu, Cd, and Pb, in order to remove high salt matrices. The preconcentration system accepts digests with acid concentrations equivalent to 0.8-1.4 % HN03, neutralizes them, and loads them onto a macroporous imlnodlacetate resin. The alkali and alkaline-earth metals, along with deleterious anions such as chloride, are washed off the resin before and analyte metals are eluted with nitric acid. A total of 13 isotopes of the analytes are measured. An examination of the apparent concentration efficiencies, as well as the behavior of two Internal standard elements added to the eluate stream, Indicates that the elution front matrix enhances the inductively coupled plasma mass spectrometry response of the analytes. Investigation of the nature of the blank signals suggests that the detection limits of several of the isotopes could benefit by much larger preconcentration factors, while those of vanadium, copper, cadmium, and lead are currently limited by reagent purity and laboratory contamination. The preconcentration system is evaluated on several simple synthetic matrices, as well as on synthetic seawater and three wastewaters. Digestion of samples containing natural organic chelators is required. © 1990, American Chemical Society. All rights reserved.
引用
收藏
页码:857 / 864
页数:8
相关论文
共 25 条
[1]  
[Anonymous], 1986, TEST METHODS EVALUAT
[2]   DETERMINATION OF TRACE-METALS IN REFERENCE WATER STANDARDS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY WITH ONLINE PRECONCENTRATION [J].
BEAUCHEMIN, D ;
BERMAN, SS .
ANALYTICAL CHEMISTRY, 1989, 61 (17) :1857-1862
[3]   DETERMINATION OF TRACE-METALS IN A RIVER WATER REFERENCE MATERIAL BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
MCLAREN, JW ;
MYKYTIUK, AP ;
BERMAN, SS .
ANALYTICAL CHEMISTRY, 1987, 59 (05) :778-783
[4]   STUDY OF THE EFFECTS OF CONCOMITANT ELEMENTS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
MCLAREN, JW ;
BERMAN, SS .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1987, 42 (03) :467-490
[5]   ONLINE SEPARATION OF SILVER WITH THE CHELATE-FORMING RESIN AMBERLITE XAD-2-PAR PRIOR TO ITS DETERMINATION BY FLAME ATOMIC-ABSORPTION SPECTROMETRY [J].
BRAJTER, K ;
DABEKZLOTORZYNSKA, E .
ANALYST, 1988, 113 (10) :1571-1574
[7]   STUDY OF SOLIDS DEPOSITION ON INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY SAMPLERS AND SKIMMERS [J].
DOUGLAS, DJ ;
KERR, LA .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (06) :749-752
[8]   FLOW-INJECTION ANALYSIS UTILIZING A SPECTRALLY SEGMENTED PHOTODIODE-ARRAY INDUCTIVELY COUPLED PLASMA EMISSION SPECTROMETER .1. MICROCOLUMN PRECONCENTRATION FOR THE DETERMINATION OF MOLYBDENUM [J].
FURUTA, N ;
BRUSHWYLER, KR ;
HIEFTJE, GM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1989, 44 (04) :349-358
[9]   TRACE-ELEMENT ANALYSIS OF NATURAL-WATER SAMPLES BY NEUTRON-ACTIVATION ANALYSIS WITH CHELATING RESIN [J].
GREENBERG, RR ;
KINGSTON, HM .
ANALYTICAL CHEMISTRY, 1983, 55 (07) :1160-1165
[10]   INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRIC DETERMINATION OF TRACE-ELEMENTS IN SURFACE WATERS SUBJECT TO ACIDIC DEPOSITION [J].
HENSHAW, JM ;
HEITHMAR, EM ;
HINNERS, TA .
ANALYTICAL CHEMISTRY, 1989, 61 (04) :335-342