STEROID-TIPPED LEADS VERSUS POROUS PLATINUM PERMANENT PACEMAKER LEADS - A CONTROLLED-STUDY

被引:27
作者
WISH, M [1 ]
SWARTZ, J [1 ]
COHEN, A [1 ]
COHEN, R [1 ]
FLETCHER, R [1 ]
机构
[1] VET ADM MED CTR,DEPT MED,DIV CARDIOL,WASHINGTON,DC 20422
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1990年 / 13卷 / 12期
关键词
PACEMAKER LEADS; STEROID-TIPPED LEADS;
D O I
10.1111/j.1540-8159.1990.tb06910.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
There is little data directly comparing steroid-tipped permanent pacemaker leads to otherwise state-of-the-art porous platinum leads. Eighteen patients receiving unipolar generators capable of low voltage outputs were randomized at the time of implant to receive either steroid-tipped leads or porous platinum leads. All leads were unipolar, tined, passive fixation, and placed in the right ventricular apex or atrial appendage. This study is single center. At implant, threshold pulse width was determined at 3 voltages (2.5, 1.5, and 0.8 V). Follow-up thresholds were determined at weeks 1, 2, 3, and 4, and at 3 and 6 months. There was no difference in implant thresholds or amplitudes for sensing. By 2 weeks postimplant, lower thresholds were noted for the steroid leads, and this discrepancy grew more significant with time. There was no significant postimplant rise in threshold for steroid-tipped leads. At 6 months, the average threshold pulse width for ventricular steroid leads at 0.8 V was 0.3 +/- 0.1 msec. In contrast, five patients with standard leads did not capture at maximum pulse width at 0.8 V (P < 0.0001). There was no significant difference in the amplitude of the chronic atrial electrogram. This study shows steroidtipped leads to offer a significant advantage in reducing thresholds early postimplant and chronically.
引用
收藏
页码:1887 / 1890
页数:4
相关论文
共 10 条
  • [1] BENDITT DG, 1985, PACEMAKER LEADS, P323
  • [2] HAYES DL, 1985, PACEMAKER LEADS, P317
  • [3] KRUSE I, 1985, PACEMAKER LEADS, P255
  • [4] LONG-TERM PERFORMANCE OF ENDOCARDIAL LEADS WITH STEROID-ELUTING ELECTRODES
    KRUSE, IM
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (06): : 1217 - 1219
  • [5] THE POROUS TITANIUM STEROID ELUTING ELECTRODE - A DOUBLE-BLIND STUDY ASSESSING THE STIMULATION THRESHOLD EFFECTS OF STEROID
    MOND, H
    STOKES, K
    HELLAND, J
    GRIGG, L
    KERTES, P
    PATE, B
    HUNT, D
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (02): : 214 - 219
  • [6] Mugica J., 1984, CLIN PROG PACING ELE, V2, P513
  • [7] CHANGES IN MYOCARDIAL THRESHOLD . PHYSIOLOGIC AND PHARMACOLOGIC FACTORS IN PATIENTS WITH IMPLANTED PACEMAKERS
    PRESTON, TA
    FLETCHER, RD
    LUCCHESI, BR
    JUDGE, RD
    [J]. AMERICAN HEART JOURNAL, 1967, 74 (02) : 235 - &
  • [8] Stokes K, 1985, MODERN CARDIAC PACIN, P33
  • [9] STOKES KB, 1983, 7TH P WORLD S CARD P, P369
  • [10] Westerholm C J, 1971, Scand J Thorac Cardiovasc Surg Suppl, V8, P1