EFFECTS OF VOLTAGE AND RESPIRATION ON IMPEDANCE IN NONTHORACOTOMY DEFIBRILLATION PATHWAYS

被引:14
作者
SWERDLOW, CD
KASS, RM
HWANG, C
GANG, E
CHEN, PS
PETER, T
机构
[1] CEDARS SINAI MED CTR,DIV CARDIOL,LOS ANGELES,CA 90048
[2] CEDARS SINAI MED CTR,DIV CARDIOTHORAC SURG,LOS ANGELES,CA
[3] DANIEL FREEMAN MEM HOSP,DIV CARDIOL,LOS ANGELES,CA
关键词
D O I
10.1016/0002-9149(94)90935-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The effects of applied voltage and phase of respiration on impedance of pathways used by implantable cardioverter-defibrillators were investigated. Patients were studied at implantation of cardioverter-defibrillators using epicardial (n = 12) or transvenous and subcutaneous (SQ) (n = 30) electrodes. Transvenous-SQ pathways were right ventricular cathode to SQ anode and coronary sinus cathode to SQ anode. Transvenous-transvenous pathways were right ventricle to coronary sinus and right ventricle to superior vena cava. Patients with nonthoracotomy electrode systems were studied at end-expiration and end-inspiration. Five shocks of 65 to 745 V (0.2 to 34 J) were given in random order in sinus rhythm. Over this range, end-expiratory impedance decreased monotonically for all pathways. This effect was greatest for transvenous-SQ pathways (13 +/- 3% to 17 +/- 4%, p <0.001), intermediate for transvenous-transvenous pathways (5 +/- 4% to 8 +/- 5%, p <0.001), and least for epicardial pathways (3 +/- 3%, p = 0.006). Paired data in inspiration and expiration showed that inspiration increased impedance in transvenous-SQ pathways (p <0.001) but not in transvenous-transvenous pathways. Further, the effects of respiration and voltage on impedance in transvenous-SQ pathways were interactive (p <0.001): Inspiration increased voltage-dependence of impedance. The magnitude of the inverse relationship between voltage and impedance depends on type of defibrillation pathway. The effect of respiration on impedance suggests that voltage-dependence of impedance is greatest in the lungs. These findings have potential relevance for intraoperative testing of cardioverter-defibrillators and selection of pathways for low-energy cardioversion.
引用
收藏
页码:688 / 692
页数:5
相关论文
共 27 条
[1]   USE OF ELECTRICAL IMPEDANCE TECHNIQUES FOR MONITORING OF RESPIRATORY PATTERN DURING ANAESTHESIA [J].
BAKER, LE ;
HILL, DW .
BRITISH JOURNAL OF ANAESTHESIA, 1969, 41 (01) :2-&
[2]   CLINICAL-EXPERIENCE WITH A TIERED-THERAPY, MULTIPROGRAMMABLE ANTIARRHYTHMIA DEVICE [J].
BARDY, GH ;
TROUTMAN, C ;
POOLE, JE ;
KUDENCHUK, PJ ;
DOLACK, GL ;
JOHNSON, G ;
HOFER, B .
CIRCULATION, 1992, 85 (05) :1689-1698
[3]   IMPLANTABLE TRANSVENOUS CARDIOVERTER-DEFIBRILLATORS [J].
BARDY, GH ;
HOFER, B ;
JOHNSON, G ;
KUDENCHUK, PJ ;
POOLE, JE ;
DOLACK, GL ;
GLEVA, M ;
MITCHELL, R ;
KELSO, D .
CIRCULATION, 1993, 87 (04) :1152-1168
[4]   STRENGTH-DURATION CURVES OF FIXED PULSE WIDTH VARIABLE TILT TRUNCATED EXPONENTIAL WAVEFORMS FOR NONTHORACOMY INTERNAL DEFIBRILLATION IN DOGS [J].
CHAPMAN, PD ;
WETHERBEE, JN ;
VETTER, JW ;
TROUP, P ;
SOUZA, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (07) :1045-1050
[5]   THE POTENTIAL GRADIENT FIELD CREATED BY EPICARDIAL DEFIBRILLATION ELECTRODES IN DOGS [J].
CHEN, PS ;
WOLF, PD ;
CLAYDON, FJ ;
DIXON, EG ;
VIDAILLET, HJ ;
DANIELEY, ND ;
PILKINGTON, TC ;
IDEKER, RE .
CIRCULATION, 1986, 74 (03) :626-636
[6]  
Dahl C F, 1976, Med Instrum, V10, P151
[7]  
DEGARVILLA L, 1981, P ASS ADV MED INSTRU, V16, P28
[8]   CANINE TRANSTHORACIC RESISTANCE [J].
EWY, GA ;
EWY, MD ;
NUTTALL, AW ;
NUTTALL, AJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1972, 32 (01) :91-&
[9]   INFLUENCE OF VENTILATION PHASE ON TRANS-THORACIC IMPEDANCE AND DEFIBRILLATION EFFECTIVENESS [J].
EWY, GA ;
HELLMAN, DA ;
MCCLUNG, S ;
TAREN, D .
CRITICAL CARE MEDICINE, 1980, 8 (03) :164-166
[10]  
EWY GA, 1978, MED INSTRUM, V12, P47