DIAPHRAGM MAGNETOMETER FOR DC MEASUREMENTS IN HIGH MAGNETIC-FIELDS

被引:14
作者
SWANSON, AG
MA, YP
BROOKS, JS
MARKIEWICZ, RM
MIURA, N
机构
[1] NORTHEASTERN UNIV,DEPT PHYS,BOSTON,MA 02115
[2] UNIV TOKYO,INST SOLID STATE PHYS,ROPPONGI,JAPAN
关键词
D O I
10.1063/1.1141452
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe the design and operation of a force magnetometer with high sensitivity which can be easily used between room and sub-kelvin temperatures. The device can accommodate a large range of sample masses (10-0.001 g) with a relative resolution of 10-5 emu. The maximum sample size is only restricted by the geometry of the cryogenic and magnetic environment. The design is particularly suited for high magnetic field measurements or where no gradient coil is available.
引用
收藏
页码:848 / 851
页数:4
相关论文
共 11 条
[1]   MAGNETIZATION OF HIGH-FIELD SUPERCONDUCTORS [J].
BEAN, CP .
REVIEWS OF MODERN PHYSICS, 1964, 36 (1P1) :31-+
[2]   SMALL SAMPLE MAGNETOMETERS FOR SIMULTANEOUS MAGNETIC AND RESISTIVE MEASUREMENTS AT LOW-TEMPERATURES AND HIGH MAGNETIC-FIELDS [J].
BROOKS, JS ;
NAUGHTON, MJ ;
MA, YP ;
CHAIKIN, PM ;
CHAMBERLIN, RV .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1987, 58 (01) :117-121
[3]   SINGLE DOMAIN MAGNETIC PARTICLES - MAGNETIC ALIGNMENT AND EVIDENCE FOR VERWEY TRANSITION FROM LOW-TEMPERATURE MEASUREMENTS [J].
DEFRANZO, A ;
GUNTHER, L ;
SWANSON, AG ;
BROOKS, JS .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1987, 26 :843-844
[4]  
MA YP, 1986, EXTENDED ABSTRACTS, P149
[5]  
MA YP, 1988, THESIS BOSTON U
[6]  
MCNIFF EJ, COMMUNICATION
[7]  
NAUGHTON MJ, 1986, THESIS BOSTON U
[8]  
OBERG E, 1951, MACHINERY HDB
[9]   NEAREST-NEIGHBOR EXCHANGE CONSTANT AND MN DISTRIBUTION IN ZN1-XMNXTE FROM HIGH-FIELD MAGNETIZATION STEP AND LOW-FIELD SUSCEPTIBILITY [J].
SHAPIRA, Y ;
FONER, S ;
BECLA, P ;
DOMINGUES, DN ;
NAUGHTON, MJ ;
BROOKS, JS .
PHYSICAL REVIEW B, 1986, 33 (01) :356-365
[10]  
SWANSON AG, 1988, B AM PHYS SOC, V33, P363