PRELIMINARY EVALUATION OF A NEW ULTRASOUND BONE DENSITOMETER

被引:104
作者
LEES, B
STEVENSON, JC
机构
[1] Wynn Institute for Metabolic Research, London
关键词
ULTRASOUND; BONE MINERAL DENSITY; OSTEOPOROSIS;
D O I
10.1007/BF01321829
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A new ultrasound bone densitometer has been developed that measures ultrasonic properties of the os calcis, namely, the speed of sound (SOS), broadband ultrasound attenuation (BUA), and a proprietary factor derived from SOS and BUA, termed ''stiffness.'' Short-term precision of ultrasound measurements was 1.4% for BUA, 0.2% for SOS, and 1.5% for stiffness in healthy women, and 1.1% for BUA, 0.1% for SOS, and 1.5% for stiffness in osteopenic women. One hundred seven women underwent measurements by ultrasound, together with dual energy X-ray absorptiometry (DXA) bone mineral density (BMD) measurements of the lumbar spine and proximal femur. Correlations between SOS, BUA, and stiffness measurements and DXA BMD measurements were all highly significant (P < 0.001) with r values varying from 0.54 to 0.67. BUA, SOS, and stiffness measurements were all significantly different between normal and osteopenic women even after adjusting for age, height, and weight (P < 0.05, P < 0.001, and P < 0.01, respectively). These results demonstrate that this ultrasound system measures ultrasonic properties of the os calcis with good precision, the measurements correlate moderately well with DXA BMD measurements and they can differentiate between normals and those with osteopenia.
引用
收藏
页码:149 / 152
页数:4
相关论文
共 17 条
  • [1] ULTRASOUND ATTENUATION OF THE CALCANEUS - A SENSITIVE AND SPECIFIC DISCRIMINATOR OF OSTEOPENIA IN POSTMENOPAUSAL WOMEN
    AGREN, M
    KARELLAS, A
    LEAHEY, D
    MARKS, S
    BARAN, D
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1991, 48 (04) : 240 - 244
  • [2] Baran D T, 1991, Osteoporos Int, V1, P110, DOI 10.1007/BF01880452
  • [3] ULTRASOUND ATTENUATION OF THE OS CALCIS IN WOMEN WITH OSTEOPOROSIS AND HIP-FRACTURES
    BARAN, DT
    KELLY, AM
    KARELLAS, A
    GIONET, M
    PRICE, M
    LEAHEY, D
    STEUTERMAN, S
    MCSHERRY, B
    ROCHE, J
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1988, 43 (03) : 138 - 142
  • [4] CAMPOREALE A, 1990, OSTEOPOROSIS 90, P670
  • [5] OSTEOPOROTIC BONE FRAGILITY - DETECTION BY ULTRASOUND TRANSMISSION VELOCITY
    HEANEY, RP
    AVIOLI, LV
    CHESNUT, CH
    LAPPE, J
    RECKER, RR
    BRANDENBURGER, GH
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1989, 261 (20): : 2986 - 2990
  • [6] EFFECTIVE DOSE VALUES IN BONE-MINERAL MEASUREMENTS BY PHOTON-ABSORPTIOMETRY AND COMPUTED-TOMOGRAPHY
    KALENDER, WA
    [J]. OSTEOPOROSIS INTERNATIONAL, 1992, 2 (02) : 82 - 87
  • [7] Langton C M, 1984, Eng Med, V13, P89, DOI 10.1243/EMED_JOUR_1984_013_022_02
  • [8] AN EVALUATION OF DUAL-ENERGY X-RAY ABSORPTIOMETRY AND COMPARISON WITH DUAL-PHOTON ABSORPTIOMETRY
    LEES, B
    STEVENSON, JC
    [J]. OSTEOPOROSIS INTERNATIONAL, 1992, 2 (03) : 146 - 152
  • [9] PERFORMANCE EVALUATION OF A DUAL-ENERGY X-RAY BONE DENSITOMETER
    MAZESS, R
    COLLICK, B
    TREMPE, J
    BARDEN, H
    HANSON, J
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1989, 44 (03) : 228 - 232
  • [10] ASSESSMENT OF BROAD-BAND ULTRASOUND ATTENUATION IN THE OS CALCIS INVITRO
    MCCLOSKEY, EV
    MURRAY, SA
    CHARLESWORTH, D
    MILLER, C
    FORDHAM, J
    CLIFFORD, K
    ATKINS, R
    KANIS, JA
    [J]. CLINICAL SCIENCE, 1990, 78 (02) : 221 - 225