BIOMECHANICAL ANALYSIS OF 3 SURGICAL APPROACHES FOR LUMBAR BURST FRACTURES USING SHORT-SEGMENT INSTRUMENTATION

被引:117
作者
GURWITZ, GS
DAWSON, JM
MCNAMARA, MJ
FEDERSPIEL, CF
SPENGLER, DM
机构
[1] Department of Orthopaedics and Rehabilitation, Vanderbilt University School of Medicine, Nashville, TN
[2] Division of Biostatistics, Department of Preventive Medicine, Vanderbilt University School of Medicine, Nashville, TN
关键词
LUMBAR SPINE; BIOMECHANICAL TESTING; ANTERIOR INSTRUMENTATION; POSTERIOR INSTRUMENTATION; CORPECTOMY; BURST FRACTURE;
D O I
10.1097/00007632-199306150-00005
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Burst fractures of the lumbar spine that are located below the thoracolumbar junction present a challenge when operative management is indicated. Short-segment instrumentation offers the advantage of incorporating fewer motion segments in the fusion, but may not provide adequate long-term stabilization. The goal of this study was to assess the axial stiffness and torsional rigidity of several short-segment instrumentation procedures. Compressive axial stiffness and torsional rigidity were measured in six intact porcine lumbar spines (Ll-L5). A corpectomy was performed to simulate a burst fracture injury and decompression. Posterior instrumentation, posterior instrumentation with an anterior strut (a wood block), and anterior instrumentation with an anterior strut one level above and one level below the fracture site were applied as treatment strategies. VSP plates (Acromed, Cleveland, OH) for posterior instrumentation and the Kaneda system (Acromed, Cleveland, OH) for anterior instrumentation were used. Load-displacement and torque-angle plots were generated and used to calculate 144 estimates of axial stiffness and 144 estimates of torsional rigidity for these constructs. These analyses showed that, in comparison with the intact spine, posterior instrumentation alone was an average of 76% less stiff axially, posterior instrumentation with an anterior strut was 3% more stiff (not significantly different from intact), and anterior instrumentation with an anterior strut was 15% more stiff. Posterior instrumentation alone was an average of 30% less rigid in torsion, posterior instrumentation with an anterior strut was 26% less rigid, and anterior instrumentation with an anterior strut was 24% less rigid than the intact spine. These results suggest that short-segment posterior instrumentation alone does not restore the stiffness or rigidity of the injured spine to the stiffness and rigidity of the intact spine. Further, these results suggest that the reconstruction of the anterior column with a rigid strut is an important adjunct to posterior or anterior instrumentation.
引用
收藏
页码:977 / 982
页数:6
相关论文
共 21 条
[1]  
Ashman R., Biomechanical analysis of pedicle screw instrumentation systems in a corpectomy model, Spine, 14, 12, pp. 1398-1405, (1989)
[2]  
Been H., Anterior decompression and stabilization of the thoracolumbar burst fracture by use of the Slot-Zielke device, Spine, 16, pp. 70-77, (1991)
[3]  
Cochran T., Irstam L., Nachemson A., Long-term anatomic and functional changes in patients with adolescent idiopathic scoliosis treated by Harrington rod fusion, Spine, 8, pp. 576-584, (1983)
[4]  
Denis F., The three-column spine and its significance in the classification of acute thoracolumbar spine injuries, Spine, 8, pp. 817-831, (1983)
[5]  
Dunn H., Anterior stabilization of thoracolumbar injuries, CORR, 189, pp. 116-124, (1989)
[6]  
Gaines R., Carson W., Satterlee C., Groh G., Experimental evaluation of seven different spinal fracture internal fixation devices using nonfailure stability testing the sharing and unstable mechanism concepts, Spine, 16, pp. 902-909, (1991)
[7]  
Garfin S., Mowery C., Guerra J., Confirmation of the posterolateral technique to decompress and fuse thoracolumbar spine burst fractures, Spine, 10, pp. 218-231, (1985)
[8]  
Gertzbein S., Brown C., Jacobs R., Decompression and circumferential stabilization of unstable spinal fractures, Spine, 13, pp. 892-895, (1988)
[9]  
Gurr K., McAfee P., Biomechanical analysis of anterior and posterior instrumentation systems after corpectomy calf spine model, J Bone Joint Surg [Am], 78A, pp. 1182-1191, (1988)
[10]  
Harrington P., Treatment of scoliosis, correction and internal fixation by spine instrumentation, J Bone Joint Surg [Am], 44A, (1962)