PERFORMANCE-CHARACTERISTICS OF AN ULTRASONIC NEBULIZER COUPLED TO A 40.68 MHZ INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROMETER

被引:27
作者
BRENNER, IB [1 ]
BREMIER, P [1 ]
LEMARCHAND, A [1 ]
机构
[1] JOBIN YVON,F-91163 LONGJUMEAU,FRANCE
关键词
SAMPLE INTRODUCTION; INDUCTIVELY COUPLED PLASMA; TRACE ELEMENT ANALYSIS OF WATER; ULTRASONIC NEBULIZATION;
D O I
10.1039/ja9920700819
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Analytical performance characteristics [limits of detection (LODs), memory effects, long and short-time reproducibility, effect of high salt content and accuracy] of an air-cooled ultrasonic nebulizer coupled to a 40.68 MHz r.f. inductively coupled plasma (ICP) were evaluated. Approximately 10-fold enhancements in the LODs, only slightly affected by ICP power, were observed with a simultaneous, sequential polychromator. Limits of detection were somewhat better than those quoted in the literature for 27 MHz r.f. ICP generators. The present configuration allowed the determination of trace elements at the mug dm-3 and sub-mug dm-3 concentration level in pristine and waste waters. The use of a Trassy-Mermet sheath gas assembly and a 3 mm injector tube allowed the direct analysis of saline solutions containing trace elements at the mug dm-3 level using matrix-matched calibration protocols. With this configuration, memory effects were minimum and the long- and short-term variations could be compensated for with the use of internal references. As a result relative standard deviations are improved and the need for frequent system readjustment and recalibration was eliminated. A comparison of the data for standard reference and spiked materials indicated that the accuracy was satisfactory.
引用
收藏
页码:819 / 824
页数:6
相关论文
共 23 条
[1]  
[Anonymous], TRACE ELEMENTS WATER
[2]   SIMULTANEOUS DETERMINATION OF 5 TRACE-METALS IN SEAWATER BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY WITH ULTRASONIC NEBULIZATION [J].
BERMAN, SS ;
MCLAREN, JW ;
WILLIE, SN .
ANALYTICAL CHEMISTRY, 1980, 52 (03) :488-492
[3]   STUDIES OF AN INDUCTIVELY-COUPLED HIGH-FREQUENCY ARGON PLASMA FOR OPTICAL EMISSION SPECTROMETRY .2. COMPROMISE CONDITIONS FOR SIMULTANEOUS MULTIELEMENT ANALYSIS [J].
BOUMANS, PWJM ;
DEBOER, FJ .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1975, 30 (09) :309-334
[4]  
BRENNER IB, 1986, ICP INFORM NEWSL, V12, P243
[5]  
BRENNER IB, 1984, ICP INFORMATION NEWS, V10, P451
[6]  
BRENNER IB, 1989, 1989 PITTSB C ATL
[7]  
BRENNER IB, 1991, 27 C SPECTR INT BERG
[8]   INFLUENCE OF THE GENERATOR FREQUENCY ON THE SPECTRAL CHARACTERISTICS OF INDUCTIVELY COUPLED PLASMA [J].
CAPELLE, B ;
MERMET, JM ;
ROBIN, J .
APPLIED SPECTROSCOPY, 1982, 36 (02) :102-106
[9]  
CHAN SK, 1989, 21ST ANN C CAN MIN A
[10]  
CHAN SK, 1990, ANAL WATER WASTES US