SURFACE COIL IMAGING OF RAT SPINE AT 7.0-T

被引:11
作者
BANSON, ML [1 ]
COFER, GP [1 ]
HEDLUND, LW [1 ]
JOHNSON, GA [1 ]
机构
[1] DUKE UNIV,MED CTR,CTR INVIVO MICROSCOPY,BOX 3302,DURHAM,NC 27710
关键词
MAGNETIC RESONANCE MICROSCOPY; SPINAL CORD; SURFACE COILS;
D O I
10.1016/0730-725X(92)90447-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An inductively coupled surface coil for imaging the rat spine at 7 T is described. This planar circular probe was made from microwave substrate to limit the size of the coil and to minimize the magnetic susceptibility. The surface coil was used as a single transmit/receive coil and as a receive-only coil with a birdcage body coil for excitation. The signal-to-noise ratio (SNR) of the probe was compared to a 5-cm birdcage coil and exceeded the birdcage coil's SNR by three to six times at superficial structures. The main advantages of the probe are an improved SNR for superficial structures and a simple design and use. Images with 50 x 50 x 500 mum voxels were obtained of the rat spine with excellent anatomical detail.
引用
收藏
页码:929 / 934
页数:6
相关论文
共 12 条
[1]  
Fisher, Barker, Amparo, Brandt, Brant-Zawadzki, Hricak, Higgins, MR imaging using specialized coils, Radiology, 157, pp. 443-447, (1985)
[2]  
Edelman, McFarland, Stark, Ferrucci, Simeone, Wismer, White, Rosen, Brady, Surface coil MR imaging of abdominal viscera: Part 1 — Theory, technique, and initial results, Radiology, 157, pp. 425-430, (1985)
[3]  
Reiman, Heiken, Totty, Lee, Clinical MR imaging with a Helmholtz-type surface coil, Radiology, 169, pp. 564-566, (1988)
[4]  
Schenck, Foster, Henkes, Adams, Hayes, Hart, Edelstein, Bottomley, Wehrli, High-field surface-coil MR imaging of localized anatomy, AJNR, 6, pp. 181-186, (1985)
[5]  
Rudin, MR microscopy on rats in vivo at 4.7 T using surface coils, Magn. Reson. Med., 5, pp. 443-448, (1987)
[6]  
Hyde, Kneeland, High-resolution methods using local coils, Biomedical Magnetic Resonance Imaging, pp. 189-223, (1988)
[7]  
Schenck, Hart, Foster, Edelstein, Hussain, High resolution magnetic resonance imaging using surface coils, Magnetic Resonance Annual, pp. 123-160, (1986)
[8]  
Crowley, Evelhoch, Ackerman, The surface-coil NMR receiver in the presence of homogeneous B<sub>1</sub> excitation, J. Magn. Reson., 64, pp. 20-31, (1985)
[9]  
Hedlund, Johnson, Mills, Magnetic resonance microscopy of the rat thorax and abdomen, Invest. Radiol., 21, 11, pp. 843-846, (1986)
[10]  
Hedlund, Deitz, Nassar, Et al., A ventilator for magnetic resonance imaging, Invest. Radiol., 21, pp. 18-23, (1986)