A PHANTOM FOR QUANTITATIVE ULTRASOUND OF TRABECULAR BONE

被引:52
作者
CLARKE, AJ
EVANS, JA
TRUSCOTT, JG
MILNER, R
SMITH, MA
机构
[1] Centre for Bone and Body Composition Res., Leeds Univ.
关键词
D O I
10.1088/0031-9155/39/10/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The propagation mechanisms of ultrasound in trabecular bone are poorly understood and have been the subject of extended debate; also, the reproducibility of ultrasonic measurements on bone in vivo using commercial ultrasound heel-scanning devices is such that the interpretation of the obtained data is difficult. In this paper we describe recent developments in the production of a bone-mimicking material which is well suited to the task of routine monitoring of commercial ultrasound bone scanners. The material, based on a standard epoxy resin is fabricated to a pre-determined porosity value by the inclusion of a marrow-mimicking material thereby introducing a known and controlled mean pore size. Measurements of the velocity and attenuation of the material have been performed over a range of porosity values from 10% to 80% in the frequency range 500-900 kHz; also, broadband ultrasonic attenuation (BUA) values have been obtained from commercial equipment. The material displays velocities in the range 1844-3118 m s(-1) and attenuation ranging from 7.0 to 17.7 dB cm(-1) at 500 kHz.
引用
收藏
页码:1677 / 1687
页数:11
相关论文
共 16 条
  • [1] MEASUREMENT OF THE VELOCITY OF ULTRASOUND IN THE HUMAN FEMUR INVIVO
    ANDRE, MP
    CRAVEN, JD
    GREENFIELD, MA
    [J]. MEDICAL PHYSICS, 1980, 7 (04) : 324 - 330
  • [4] ULTRASONIC-ATTENUATION AND VELOCITY IN BONE
    EVANS, JA
    TAVAKOLI, MB
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (10) : 1387 - 1396
  • [5] Langton C M, 1984, Eng Med, V13, P89, DOI 10.1243/EMED_JOUR_1984_013_022_02
  • [6] A CONTACT METHOD FOR THE ASSESSMENT OF ULTRASONIC VELOCITY AND BROAD-BAND ATTENUATION IN CORTICAL AND CANCELLOUS BONE
    LANGTON, CM
    ALI, AV
    RIGGS, CM
    EVANS, GP
    BONFIELD, W
    [J]. CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1990, 11 (03): : 243 - 249
  • [7] LANGTON CM, 1984, THESIS U HULL
  • [8] ASSESSMENT OF BROAD-BAND ULTRASOUND ATTENUATION IN THE OS CALCIS INVITRO
    MCCLOSKEY, EV
    MURRAY, SA
    CHARLESWORTH, D
    MILLER, C
    FORDHAM, J
    CLIFFORD, K
    ATKINS, R
    KANIS, JA
    [J]. CLINICAL SCIENCE, 1990, 78 (02) : 221 - 225
  • [9] MCKELVIE ML, 1987, IOP SHORT M SERIES, V6, P1
  • [10] ULTRASONIC VELOCITY-MEASUREMENTS THROUGH THE CALCANEUS - WHICH VELOCITY SHOULD BE MEASURED
    MILLER, CG
    HERD, RJM
    RAMALINGAM, T
    FOGELMAN, I
    BLAKE, GM
    [J]. OSTEOPOROSIS INTERNATIONAL, 1993, 3 (01) : 31 - 35