STRUT FRACTURE WITH BJORK-SHILEY 70-PERCENT CONVEXO-CONCAVE VALVE - AN INTERNATIONAL MULTIINSTITUTIONAL FOLLOW-UP-STUDY

被引:44
作者
ERICSSON, A [1 ]
LINDBLOM, D [1 ]
SEMB, G [1 ]
HUYSMANS, HA [1 ]
THULIN, LI [1 ]
SCULLY, HE [1 ]
BENNETT, JG [1 ]
OSTERMEYER, J [1 ]
GRUNKEMEIER, GL [1 ]
机构
[1] KAROLINSKA HOSP,DEPT THORAC SURG,S-10401 STOCKHOLM 60,SWEDEN
关键词
PROSTHETIC HEART VALVES; MECHANICAL FAILURE; STRUT FRACTURE; BJORK-SHILEY HEART VALVE;
D O I
10.1016/1010-7940(92)90169-X
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Between 1980 and 1983, 831 Bjork-Shiley 70-degrees convexo-concave prosthetic heart valves were implanted at five institutions in Sweden, Germany, The Netherlands, and Canada. As of January 1991, there were 34 outlet strut fractures occurring from 0.2 to 10.1 years (median = 4.6 years) after implantation. In addition, there were 28 sudden, unexplained deaths. The mortality after strut fracture was 84%. The mortality after emergency valve replacement for strut fracture was 50%. The 10-year actuarial fracture rate(standard error) was 10.5 (2.4)% for large (29-33 mm) valves vs. 3.3 (1.2)% for 21-27 mm valves (P < 0.001). Within valve size groups, fracture rates for aortic and mitral valves were similar. Cox regression analysis found only valve size to be significantly associated with strut fracture. There is a further subgrouping of the valves according to the manufacturer: group I are the earlier large size (29-33 mm) valves; group II are the later large valves; group III are the small size (21-27 mm) valves. The risk of strut fracture was highest in group I (12.3% at 10 years) with an approximatively constant hazard (1.4% per year). A comparison was made with a statistical model incorporating all cases reported to the manufacturer. This model estimates fracture rates approximately 63%-73% of those found in the present study. These findings lead us to recommend that group I patients should be considered for elective reoperation on an individual basis, giving careful attention to risk factors and contraindications.
引用
收藏
页码:339 / 346
页数:8
相关论文
共 24 条
[1]  
Berry B.E., Sheedy P.F., McGoon D.C., Diagnosis and management of aortic poppet embolism, Ann Thorac Surg, 17, (1974)
[2]  
Blackstone E.H., Kirklin J.W., Pluth J.R., Turner M.E., Grant G.V.S., The performance of the Braunwald-Cutter aortic prosthetic valve, Ann Thorac Surg, 23, pp. 302-318, (1977)
[3]  
Blackstone E.H., Kirklin J.W., Death and other time-related events after valve replacement, Circulation, 72, pp. 753-767, (1985)
[4]  
Bonnabeau R.C., Lillehei C.W., Mechanical “ball failure” in Starr-Edwards prosthetic valves. A report of three cases, J Thorac Cardiovasc Surg, 56, pp. 258-264, (1968)
[5]  
Brodin L.A., Ekestrom S., Forssell G., Lindblom D., Pehrsson S.K., Strut fracture of a Björk-Shiley tilting disc valve diagnosed by echocardiography - a case report, Eur Heart J, 9, pp. 191-193, (1988)
[6]  
Clark R.E., Edmunds L.H., Cohn L.H., Miller D.C., Weisel R.D., Guidelines for reporting morbidity and mortality after cardiac valvular operations, Eur J Cardio-thorac Surg, 2, pp. 293-295, (1988)
[7]  
Cox D.R., Regression methods and life tables, J Royal Stat Soc, Series B, 34, pp. 187-220, (1972)
[8]  
Cutler S.J., Ederer F., Maximum utilization of the life table method in analyzing survival, J Chron Dis, 8, pp. 699-712, (1958)
[9]  
Edmunds L.H., Clark R.E., Cohn L.H., Miller D.C., Weisel R.D., Guidelines for reporting morbidity and mortality after cardiac valvular operations, J Thorac Cardiovasc Surg, 96, pp. 351-353, (1988)
[10]  
Grunkemeier G.L., Thomas D.R., Starr A., Statistical considerations in the analysis and reporting of time-related events, Am J Cardiol, 39, pp. 257-258, (1977)