THE CDRH HELIX - AN INVIVO EVALUATION

被引:6
作者
ANHALT, D [1 ]
HYNYNEN, K [1 ]
DEYOUNG, D [1 ]
SHIMM, D [1 ]
KUNDRAT, M [1 ]
CETAS, T [1 ]
机构
[1] UNIV ARIZONA,DEPT ANIM RESOURCES,TUCSON,AZ 85724
关键词
Helix; Hyperthermia; Measurements; Power absorption; Temperature;
D O I
10.3109/02656739009140819
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The Helix is an electromagnetic heating device used to induce regional/systemic hyperthermia for cancer therapy. It is a resonant device operating at about 82 MHz with an aperture size of 60 cm × 40 cm (elliptical) × 40 cm long. The Helix deposits power in tissues (or phantoms) by producing a predominantly axial electric field within its radiating aperture. Five pig experiments were performed to provide in vivo verification of specific absorption rate (SAR) measurements and electric field measurements which were obtained earlier in tissue-equivalent phantom and 0.9% saline, respectively. In addition to verifying the power deposition patterns found in phantoms, the pig experiments provided valuable insight into the capabilities and limitations of electromagnetic regional heating. For example, a kidney with limited blood flow, simulating a necrotic tumor, heated very well-although the highest temperature was not always measured there. Also, fat heating may be a problem, since excessive temperatures in the fat were observed in approximately 20% of the heatings. This paper compares the in vivo temperature measurements in pigs with SARs and electric field measurements obtained in phantoms, and also provides a brief overview of results of the Helix in clinical situations. © 1990 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.
引用
收藏
页码:241 / 252
页数:12
相关论文
共 16 条
[1]   CDRH RF PHANTOM FOR HYPERTHERMIA SYSTEMS EVALUATIONS [J].
ALLEN, S ;
KANTOR, G ;
BASSEN, H ;
RUGGERA, P .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1988, 4 (01) :17-23
[2]  
BUECHLER DN, 1987, N AM HYPERTHERMIA GR, P65
[3]   DESIGN OF A CLINICAL DEEP-BODY HYPERTHERMIA SYSTEM BASED ON THE COAXIAL TEM APPLICATOR [J].
DELEEUW, AAC ;
LAGENDIJK, JJW .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1987, 3 (05) :413-421
[4]  
DEYOUNG DW, 1987, 9TH P IEEE ANN C MED, P994
[5]  
GERNER EW, 1986, ANN CLIN C CANCER, V28, P379
[6]  
GOPAL MK, 1987, 9TH IEEE ANN C ENG M, P1329
[7]  
HAHN GM, 1979, CANCER RES, V39, P2264
[8]  
LAGENDIJK JJW, 1983, J MICROWAVE POWER EE, V18, P367
[9]  
OLESON J R, 1986, International Journal of Hyperthermia, V2, P327, DOI 10.3109/02656738609004963
[10]   REGIONAL HYPERTHERMIA - A FEASIBILITY ANALYSIS [J].
PILEPICH, MV ;
MYERSON, RJ ;
EMAMI, BN ;
PEREZ, CA ;
LEYBOVICH, L ;
VONGERICHTEN, D .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1987, 3 (04) :347-351