COMPARISON OF THE E-TEST TO AGAR DILUTION, BROTH MICRODILUTION, AND AGAR DIFFUSION SUSCEPTIBILITY TESTING TECHNIQUES BY USING A SPECIAL CHALLENGE SET OF BACTERIA

被引:245
作者
BAKER, CN [1 ]
STOCKER, SA [1 ]
CULVER, DH [1 ]
THORNSBERRY, C [1 ]
机构
[1] CTR DIS CONTROL, CTR INFECT DIS, STAT & INFORMAT SYST BRANCH, HOSP INFECT PROGRAM, ATLANTA, GA 30333 USA
关键词
D O I
10.1128/JCM.29.3.533-538.1991
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
The E Test (AB Biodisk, Solna, Sweden) is a new method for performing antimicrobial susceptibility tests. It consists of an impervious carrier (5- by 50-mm strip) with a predefined antimicrobic gradient which is placed on an inoculated agar plate and processed like a disk diffusion test. Resuls are generated directly as MICs from a continuous concentration gradient covering 15 twofold dilutions, and MICs are read where the edge of the inhibition zone intersects the strip. We compared the E Test with disk diffusion, broth microdilution, and agar dilution tests by using a challenge set of 195 gram-positive and gram-negative bacteria for 14 antimicrobial agents. Also, disk diffusion, broth microdilution, and agar dilution tests were compared with each other. All test method comparisons gave > 94% agreement for the category of susceptibility. The E Test category agreement with disk diffusion and broth microdilution was 95.1%, and with agar dilution it was 95.2%. The E Test results were as reliable as the results obtained by the standard antimicrobial susceptibility testing methods.
引用
收藏
页码:533 / 538
页数:6
相关论文
共 11 条