Dual-beam, light-scattering interferometry is introduced for simuitaneous measurement of droplet-size and droplet-velocity distributions. The technique was tested in the characterization of tertiary aerosols produced by five nebulization systems (a pneumatic, an ultrasonic, and three frit-type nebulizers) used in inductively coupled plasma spectrometry. in general, the Sauter mean diameter (D3,2) of tertiary aerosol produced by frit-type nebulizer was smaller than those of ultrasorric and pneumatic nebulizers. At higher injector gas flow, D3,2 was reduced for the pneumatic nebulizer but was increased for the ultrasonic nebulizer. At the same injector gas flow rate, droplets move slightly faster if helium is used as injector gas instead of argon. For all practical purposes, the D3,2 values measured for tertiary aerosol with dual-beam, light-scattering interferometry are comparable to the same data obtained by laser Fraunhofer diffraction. Compared to laser Fraunhofer diffraction, the proposed technique possesses two unique advantages. First, droplet velocity distribution can be measured. Second, spatially resolved information on droplet-size and velocity distributions may be obtained directly and rapidly. © 1990, American Chemical Society. All rights reserved.
引用
收藏
页码:390 / 394
页数:5
相关论文
共 15 条
[1]   METHOD FOR MEASURING THE SIZE AND VELOCITY OF SPHERES BY DUAL-BEAM LIGHT-SCATTER INTERFEROMETRY [J].
BACHALO, WD .
APPLIED OPTICS, 1980, 19 (03) :363-370
[2]   PHASE DOPPLER SPRAY ANALYZER FOR SIMULTANEOUS MEASUREMENTS OF DROP SIZE AND VELOCITY DISTRIBUTIONS [J].
BACHALO, WD ;
HOUSER, MJ .
OPTICAL ENGINEERING, 1984, 23 (05) :583-590
[3]  
BROWNER RF, 1987, INDUCTIVELY COUPLE 2
[4]   EMPIRICAL-MODEL FOR ESTIMATING DROP SIZE DISTRIBUTIONS OF AEROSOLS GENERATED BY INDUCTIVELY COUPLED PLASMA NEBULIZERS [J].
CANALS, A ;
WAGNER, J ;
BROWNER, RF ;
HERNANDIS, V .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1988, 43 (9-11) :1321-1335
[5]   CHARACTERIZATION OF AN ANNULAR HELIUM INDUCTIVELY COUPLED PLASMA GENERATED IN A LOW-GAS-FLOW TORCH [J].
CHAN, SK ;
MONTASER, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1987, 42 (04) :591-597
[6]   THIMBLE GLASS FRIT NEBULIZER FOR ATOMIC SPECTROMETRY [J].
CLIFFORD, RH ;
MONTASER, A ;
SINEX, SA ;
CAPAR, SG .
ANALYTICAL CHEMISTRY, 1989, 61 (24) :2777-2784
[7]   REPRESENTATION OF AVERAGE DROP SIZES IN SPRAYS [J].
DODGE, LG .
JOURNAL OF PROPULSION AND POWER, 1988, 4 (06) :490-496
[8]   MEASUREMENT OF PARTICLE-SIZE, NUMBER DENSITY, AND VELOCITY USING A LASER INTERFEROMETER [J].
FARMER, WM .
APPLIED OPTICS, 1972, 11 (11) :2603-&
[9]  
FARMER WM, 1973, THESIS U TENNESSEE
[10]  
GUSTAVSSON A, 1987, INDUCTIVELY COUPLED