KT/V AND PROTEIN CATABOLIC RATE DETERMINATION FROM SERIAL UREA MEASUREMENT IN THE DIALYSATE EFFLUENT STREAM

被引:21
作者
GARRED, LJ
DIGIUSEPPE, B
CHAND, W
MCCREADY, W
CANAUD, B
机构
[1] MCKELLAR HOSP, THUNDER BAY, ONTARIO, CANADA
[2] LAPEYRONIE HOSP, MONTPELLIER, FRANCE
关键词
HEMODIALYSIS; PROTEIN CATABOLIC RATE; KT/V; MATHEMATICAL MODELING; UREA KINETIC MODELING; SPENT DIALYSATE;
D O I
10.1111/j.1525-1594.1992.tb00305.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A bloodless technique of evaluating protein catabolic rate (PCR) and KT/V (K, clearance; T, dialysis time; V, urea distribution volume) in hemodialysis patients is presented based on serial measurement of urea in the dialysate effluent stream. PCR follows from equating urea generation and urea removal over a 7 day cycle, changes in body stores being comparatively negligible: PCR = 0.026 [U1 + U2 + U3]/BW(dry) + 0.17, where U(i) is the amount of urea in mmol appearing in the dialysate for each session in the 7 day period. KT/V is obtained from the slope of the natural logarithm of spent dialysate urea concentration-time plot: KT/V = [- slope . T + 3 - DELTA-BW/BW(dry)]/[1 - 0.01786 . T(hr)], where DELTA-BW - amount ultrafiltered in liters. The dialysate-based approach was validated and compared with conventional urea kinetic modeling (UKM) for 17 patients studied for three consecutive dialyses. The dialysate-based and UKM values of PCR agreed well when in vivo clearance values based on total dialysate collection were used for UKM. KT/V values agreed poorly on a session-by-session basis but were nearly equivalent when averaged for the three dialyses of the week. These findings lay the foundation for UKM automation with a urea sensor in the effluent dialysate stream.
引用
收藏
页码:248 / 255
页数:8
相关论文
共 11 条
  • [1] NITROGEN-BALANCE DURING INTERMITTENT DIALYSIS THERAPY OF UREMIA
    BORAH, MF
    SCHOENFELD, PY
    GOTCH, FA
    SARGENT, JA
    WOLFSON, M
    HUMPHREYS, MH
    [J]. KIDNEY INTERNATIONAL, 1978, 14 (05) : 491 - 500
  • [2] DAUGIRDAS JT, 1989, INT J ARTIF ORGANS, V12, P411
  • [3] DAUGIRDAS JT, 1989, INT J ARTIF ORGANS, V12, P420
  • [4] GARRED L J, 1988, Blood Purification, V6, P372
  • [5] UREA KINETIC MODELING BY PARTIAL DIALYSATE COLLECTION
    GARRED, LJ
    RITTAU, M
    MCCREADY, W
    CANAUD, B
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1989, 12 (02) : 96 - 102
  • [6] GARRED LJ, 1989, ASAIO ABSTR, V7, P280
  • [7] Gotch F A, 1974, Proc Clin Dial Transplant Forum, P27
  • [8] A MECHANISTIC ANALYSIS OF THE NATIONAL COOPERATIVE DIALYSIS STUDY (NCDS)
    GOTCH, FA
    SARGENT, JA
    [J]. KIDNEY INTERNATIONAL, 1985, 28 (03) : 526 - 534
  • [9] JINDAL KK, 1987, T AM SOC ART INT ORG, V33, P286
  • [10] LOWRIE EG, 1983, KIDNEY INT, pS113