MECHANICAL-PROPERTIES OF XENOGENEIC SMALL-INTESTINAL SUBMUCOSA WHEN USED AS AN AORTIC GRAFT IN THE DOG

被引:104
作者
HILES, MC
BADYLAK, SF
LANTZ, GC
KOKINI, K
GEDDES, LA
MORFF, RJ
机构
[1] PURDUE UNIV,SCH MECH ENGN,W LAFAYETTE,IN 47907
[2] SENMED MED VENTURES,CINCINNATI,OH
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 1995年 / 29卷 / 07期
关键词
D O I
10.1002/jbm.820290714
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Small-intestinal submucosa (SIS) has been shown to induce tissue remodeling in vivo when used as a vascular graft. The present study investigated the physical and mechanical properties of remodeled aortic grafts derived from xenogeneic SIS material. Eight infrarenal aortic grafts were implanted in mongrel dogs. The grafts were explanted at 1 or 2 months and tested for compliance and hoop mechanical properties. The morphologic changes within the grafts were also characterized. The remodeling process produced graft structures which were significantly stronger than both the normal artery (P = .012) and the original SIS graft (P = .0001), and the compliance of these structures was one third that of normal artery and similar to the original SIS grafts. The remodeled grafts were >10 times the thickness of the implanted SIS. Immunohistochemical analysis of remodeled tissues suggest that the SIS material was degraded and resorbed over time. The remodeling process transformed a material which was physically and mechanically quite different from normal aorta into a blood conduit which had the physical and mechanical properties needed to function in this mammalian arterial system. (C) 1995 John Wiley and Sons, Inc.
引用
收藏
页码:883 / 891
页数:9
相关论文
共 15 条
  • [1] Abbott W.M., 1982, BIOL SYNTHETIC VASCU, P189
  • [2] ANDERSON JM, 1986, ASTM STP, V898, P180
  • [3] SMALL INTESTINAL SUBMUCOSA AS A LARGE DIAMETER VASCULAR GRAFT IN THE DOG
    BADYLAK, SF
    LANTZ, GC
    COFFEY, A
    GEDDES, LA
    [J]. JOURNAL OF SURGICAL RESEARCH, 1989, 47 (01) : 74 - 80
  • [4] COMPRESSIBILITY OF ARTERIAL WALL
    CAREW, TE
    VAISHNAV, RN
    PATEL, DJ
    [J]. CIRCULATION RESEARCH, 1968, 23 (01) : 61 - &
  • [5] FUNG YC, 1981, BIOMECHANICS
  • [6] HALPIN JC, 1984, PRIMER COMPOSITE MAT
  • [7] HARKER LA, 1982, BIOL SYNTHETIC VASCU, P101
  • [8] USE OF AVIDIN-BIOTIN-PEROXIDASE COMPLEX (ABC) IN IMMUNOPEROXIDASE TECHNIQUES - A COMPARISON BETWEEN ABC AND UNLABELED ANTIBODY (PAP) PROCEDURES
    HSU, SM
    RAINE, L
    FANGER, H
    [J]. JOURNAL OF HISTOCHEMISTRY & CYTOCHEMISTRY, 1981, 29 (04) : 577 - 580
  • [9] CONTINUOUS MEASUREMENT OF AORTIC RADIUS CHANGE INVIVO WITH AN INTRA-AORTIC ULTRASONIC CATHETER
    HUGHES, DJ
    FEARNOT, NE
    BABBS, CF
    BOURLAND, JD
    GEDDES, LA
    EGGELTON, R
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1985, 23 (03) : 197 - 202
  • [10] Lantz G C, 1990, J Invest Surg, V3, P217, DOI 10.3109/08941939009140351