THE NASPE BPEG GENERIC PACEMAKER CODE FOR ANTIBRADYARRHYTHMIA AND ADAPTIVE-RATE PACING AND ANTITACHYARRHYTHMIA DEVICES

被引:97
作者
BERNSTEIN, AD
CAMM, AJ
FLETCHER, RD
GOLD, RD
RICKARDS, AF
SMYTH, NPD
SPIELMAN, SR
SUTTON, R
机构
[1] ST GEORGES HOSP,LONDON,ENGLAND
[2] VET ADM MED CTR,WASHINGTON,DC 20422
[3] WESTMINSTER HOSP,LONDON SW1P 2AP,ENGLAND
[4] NATL HEART HOSP,LONDON W1M 8BA,ENGLAND
[5] GEORGE WASHINGTON UNIV,WASHINGTON,DC 20052
[6] ALBERT EINSTEIN MED CTR,NO DIV,PHILADELPHIA,PA 19141
[7] CORDIS CORP,MIAMI,FL
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1987年 / 10卷 / 04期
关键词
D O I
10.1111/j.1540-8159.1987.tb06035.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
引用
收藏
页码:794 / 799
页数:6
相关论文
共 7 条
[1]   REPORT OF THE NASPE MODE CODE COMMITTEE [J].
BERNSTEIN, AD ;
BROWNLEE, RR ;
FLETCHER, R ;
GOLD, RD ;
SMYTH, NPD ;
SPIELMAN, SR .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (03) :395-402
[2]  
BERNSTEIN AD, UNPUB NEW GENERIC CO
[3]  
BERNSTEIN AD, 1985, MODERN CARDIAC PACIN, P307
[4]   A NEW CODE FOR PACEMAKER OPERATING MODES [J].
BROWNLEE, RR ;
SHIMMEL, JB ;
DELMARCO, CJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1981, 4 (04) :396-399
[5]   A REVISED CODE FOR PACEMAKER IDENTIFICATION [J].
PARSONNET, V ;
FURMAN, S ;
SMYTH, NPD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1981, 4 (04) :400-403
[6]  
PARSONNET V, 1983, CIRCULATION, V68, pA227
[7]  
PARSONNET V, 1974, Circulation, V50, pA21