共 22 条
SLURRY PROCEDURES FOR THE DETERMINATION OF CADMIUM AND LEAD IN CEREAL-BASED PRODUCTS USING ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY
被引:26
作者:
VINAS, P
[1
]
CAMPILLO, N
[1
]
LOPEZGARCIA, I
[1
]
CORDOBA, MH
[1
]
机构:
[1] UNIV MURCIA,DEPT ANALYT CHEM,E-30071 MURCIA,SPAIN
来源:
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY
|
1994年
/
349卷
/
04期
关键词:
D O I:
10.1007/BF00323209
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
Simple and rapid methods for the determination of cadmium and lead in biscuits, bread and cereal-based products using the slurry-ETAAS approach are discussed. Suspensions were prepared in a 20% v/v ethanol medium. Phosphate was used as a chemical modifier for lead determination. For cadmium determination both palladium and a copper plus ammonia mixture were used. In both cases platform atomization was used and calibration was performed using aqueous standards. Results for two reference materials confirmed the reliability of the procedures. Relative standard deviations were in the range of 2.5-6.5% for cadmium and 4.5-14% for lead. Detection limits were, respectively, 0.5 and 8 ng/g.
引用
收藏
页码:306 / 310
页数:5
相关论文