The Anger camera principle has shown a practical limit of a few millimeters spatial resolution. To overcome this limit, a new gamma camera prototype has been developed, based on a position-sensitive photomultiplier tube (PSPMT) coupled with a new scintillation crystal, The Hamamatsu R2486 PSPMT is a 76-mm diameter photomultiplier tube in which the electrons produced in the conventional bi-alkali photocathode are multiplied by proximity mesh dynodes and form a charge cloud around the original coordinates of the light photon striking the photocathode. A crossed wire anode array collects the charge and detects the original position, The intrinsic spatial resolution of PSPMT is better than 0.3 mm. The scintillation crystal consists of yttrium aluminum perovskit (YAP:Ce or YAlO3:Ce). This crystal has a light efficiency of about 38% relative to NaI, no hygroscopicity and a good gamma radiation absorption, To match the characteristics of the PSPMT, a special crystal assembly was produced by the Preciosa Company, consisting of a bundle of YAP:Ce pillars where single crystals have 0.6 X 0.6 mm(2) cross section and 3 mm to 18 mm length, Preliminary results from such gamma camera prototypes show spatial resolution values ranging between 0.7 mm and 1 mm with an intrinsic detection efficiency of 37 divided by 65% for 140 keV gamma energy.
引用
收藏
页码:1474 / 1482
页数:9
相关论文
共 22 条
[2]  
BACCARO S, IN PRESS NUCL INST A
[3]  
BARONE LM, IN PRESS NUCL PHYS B
[4]  
BARRETT HH, 1981, RADIOLOGICAL IMAGING, P272
[5]  
BLAZEK K, 1993, EUROP J NUCL MED, V20, P854
[6]   SOLID ANGLE SUBTENDED BY A RECTANGULAR SLIT [J].
GOTOH, H ;
YAGI, H .
NUCLEAR INSTRUMENTS & METHODS, 1971, 96 (03) :485-&
[7]   MAXIMUM A POSTERIORI ESTIMATION OF POSITION IN SCINTILLATION CAMERAS [J].
GRAY, RM ;
MACOVSKI, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1976, 23 (01) :849-852
[8]   AN INEXPENSIVE, HARD X-RAY-IMAGING SPECTROMETER FOR USE IN X-RAY ASTRONOMY AND ATOMIC PHYSICS [J].
HAILEY, CJ ;
HARRISON, F ;
LUPTON, JH ;
ZIOCK, KP .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 276 (1-2) :340-346
[9]   POSITION-SENSITIVE PHOTOMULTIPLIER TUBES FOR SCINTILLATION IMAGING [J].
KUME, H ;
MURAMATSU, S ;
IIDA, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :359-363
[10]   NEWLY DEVELOPED PHOTOMULTIPLIER TUBES WITH POSITION SENSITIVITY CAPABILITY [J].
KUME, H ;
SUZUKI, S ;
TAKEUCHI, J ;
OBA, K .