STABILITY OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY COLUMNS PACKED WITH C-1 AND C-8 POLYSILOXANES SORBED INTO POROUS SILICA PARTICLES

被引:13
作者
ANAZAWA, TA [1 ]
CARRARO, F [1 ]
COLLINS, KE [1 ]
JARDIM, ICSF [1 ]
机构
[1] UNIV ESTADUAL CAMPINAS,INST QUIM,BR-13083970 CAMPINAS,BRAZIL
关键词
D O I
10.1016/0021-9673(94)00810-V
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Columns, packed with 10 mu m diameter porous silica particles (15 nm diameter pores) containing polydimethylsiloxane (20%, w/w) or polymethyloctylsiloxane (40%, w/w) sorbed into the pores, were tested for chromatographic stability as a function of elution volume. No significant changes in efficiency, retention factor or separation factor were observed following extended washing (up to 5000 column volumes) of either stationary phase with methanol-water mobile phases.