INTERFACE LOAD ANALYSIS FOR COMPUTER-AIDED-DESIGN OF BELOW-KNEE PROSTHETIC SOCKETS

被引:42
作者
REYNOLDS, DP [1 ]
LORD, M [1 ]
机构
[1] UNIV LONDON UNIV COLL,CTR BIOENGN,DEPT MECH ENGN,LONDON WC1E 7JE,ENGLAND
关键词
COMPUTER-AIDED DESIGN; EXTERNAL PROSTHESES; FINITE-ELEMENT MODELS; INTERFACE LOADING;
D O I
10.1007/BF02446180
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A finite-element analysis is made for the compression of soft tissues of the residual lower limb contained in a prosthetic socket. The analysis is relevant to static loading during stance in a patellar-tendon-bearing, below-knee design of socket. Values of Young's modulus are obtained experimentally for use in the model. One of the main objectives is to study the sensitivity of the loading to these values and also to other assumed conditions. Using direct pressure at the limb/socket interface and vertical stiffness as indicators, changes in material properties, socket alignment and socket rectification are investigated; assumptions about the frictional characteristic at the interface are seen to be critical in determination of load distribution. This type of analysis may provide the next stage of refinement for computer-aided socket design systems.
引用
收藏
页码:419 / 426
页数:8
相关论文
共 18 条
[1]  
CHILDRESS DL, 1987, J REHABIL RES DEV, V25, P22
[2]   DEFORMATIONS AND STRESSES IN SOFT BODY TISSUES OF A SITTING PERSON [J].
CHOW, WW ;
ODELL, EI .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1978, 100 (02) :79-87
[3]   ARTIFICIAL SENSING SKIN MIMICKING MECHANOELECTRICAL CONVERSION PROPERTIES OF HUMAN DERMIS [J].
DEROSSI, D ;
NANNINI, A ;
DOMENICI, C .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (02) :83-92
[4]  
DEWAR ME, 1986, 5 WORKD C INT SOC PR, P315
[5]   A COMPUTERIZED METHOD FOR THE DESIGN OF ABOVE-KNEE PROSTHETIC SOCKETS [J].
KROUSKOP, TA ;
MALINAUSKAS, M ;
WILLIAMS, J ;
BARRY, PA ;
MUILENBURG, AL ;
WINNINGHAM, DJ .
JOURNAL OF PROSTHETICS AND ORTHOTICS, 1989, 1 (03) :131-138
[6]  
LANIR Y, 1988, PHYS MED BIOL S1, V33, P344
[7]  
MEIER RH, 1973, ARCH PHYS MED REHAB, V54, P553
[8]  
NAKAMURA S, 1981, B PROSTHET RES, V10, P27
[9]  
Pearson J R, 1973, Bull Prosthet Res, V10, P52
[10]  
Radcliffe C, 1961, BIOMECHANICS LAB