A MULTICENTER EVALUATION OF THE BOEHRINGER-MANNHEIM ES-300 IMMUNOASSAY SYSTEM

被引:6
作者
DCOSTA, M
FELD, R
LAXDAL, V
TRUNDLE, D
COLLINSWORTH, WL
机构
[1] MT SINAI HOSP, TORONTO M5G 1X5, ONTARIO, CANADA
[2] UNIV IOWA HOSP & CLIN, IOWA CITY, IA 52242 USA
[3] ROYAL UNIV HOSP, SASKATOON, SK, CANADA
[4] MORTON PLANT HOSP, ALDER INST LAB MED, CLEARWATER, FL 34617 USA
[5] BOEHRINGER MANNHEIM CORP, INDIANAPOLIS, IN 46250 USA
关键词
IMMUNOASSAY; PRECISION; ACCURACY; LINEARITY; ANALYTE STUDIES; IMMUNOENZYME TECHNIQUES;
D O I
10.1016/0009-9120(93)90017-Z
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 [基础医学];
摘要
The ES 300 system, a fully automated multichannel immunoassay analyzer, was evaluated simultaneously for 9 weeks in four major centers. Precision, accuracy, carryover, comparison to in-house methods, and interferences were assessed for the following 17 tests: T4, T3, FT4, TSH, TBK, TBG, LH, FSH, prolactin, HCG, digoxin, cortisol, ferritin, IgE, insulin, AFP, and CEA. All centers reported good intra-lab and inter-lab precision. Accuracy was judged to be good based on correlation with in-house methods and recovery of target values in commercial and proficiency control materials. Linearity was evaluated for 14 analytes. Method biases were observed for T3 and insulin that were attributed to differences in standardization, No significant interferences from bilirubin, lipemia, and hemolysis were observed for all methods except insulin and AFP. Featuring random access capability, low daily maintenance, and high throughput, the ES 300 system performed well and met the stated claims of the manufacturer.
引用
收藏
页码:51 / 57
页数:7
相关论文
共 8 条
[1]
COLLINSWORTH W L, 1991, Clinical Chemistry, V37, P981
[2]
GLICK MR, 1981, INTERFEROGRAPHS USER
[3]
KNEDEL M, 1988, J CLIN CHEM CLIN BIO, V26, P149
[4]
KWIATKOWSKI M, 1990, CLIN CHEM, V36, P1078
[5]
MEYER HD, 1988, CLIN CHEM, V34, P113
[6]
Moore J J, 1991, J Int Fed Clin Chem, V3, P14
[7]
WEN CI, 1990, CLIN CHEM, V36, P1080
[8]
1987, TENTATIVE GUIDELINES