CONTROL OF INTRAAORTIC BALLOON PUMPING - THEORY AND GUIDELINES FOR CLINICAL-APPLICATIONS

被引:15
作者
JARON, D [1 ]
MOORE, TW [1 ]
HE, P [1 ]
机构
[1] DREXEL UNIV,DEPT ELECT & COMP ENGN,PHILADELPHIA,PA 19104
关键词
COMPUTER SIMULATION - Medical Applications - CONTROL SYSTEMS; OPTIMAL; -; Synthesis;
D O I
10.1007/BF02584236
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The effectiveness of intraaortic balloon pumping was investigated by using a lumped parameter model of the cardiovascular/assist device system. The model consists of a time-varying elastance left ventricular simulation, a 2-element windkessel arterial simulation, and an RC venous return and pulmonary simulation. The four major hemodynamic variables, stroke volume (SV), aortic mean diastolic pressure (MDP), tension time index (TTI), and aortic end diastolic pressure (EDP), were divided into two categories related to system energy supply and demand: 'external' and 'internal' variables. The effects of balloon pumping on these variables can be described by closed-form equations that yield an optimal solution. The model prediction suggests that, in the ideal case, optimization of balloon pumping calls for instantaneous inflation of the balloon to maximum volume at end systole and instantaneous complete deflation at end diastole. For finite inflation/deflation rates, the optimal time for the start of inflation is end systole.
引用
收藏
页码:155 / 175
页数:21
相关论文
共 20 条
[1]   DIASTOLIC CORONARY-ARTERY PRESSURE-FLOW RELATIONS IN DOG [J].
BELLAMY, RF .
CIRCULATION RESEARCH, 1978, 43 (01) :92-101
[2]  
BERNE RM, 1979, HDB PHYSL 2, V1, P775
[3]   FEASIBILITY OF CLOSED-LOOP CONTROL OF INTRA-AORTIC BALLOON PUMPING [J].
CLARK, JW ;
KANE, GR ;
BOURLAND, HM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1973, BM20 (06) :404-412
[4]   REGULATION OF CORONARY BLOOD-FLOW DURING INDIVIDUAL DIASTOLES IN THE DOG [J].
DOLE, WP ;
BISHOP, VS .
CIRCULATION RESEARCH, 1982, 50 (03) :377-385
[5]  
FRONEK A, 1965, CIRC RES, V14, P545
[6]   MODEL DESCRIBING RESPONSE OF CIRCULATORY-SYSTEM TO ACCELERATION STRESS [J].
GREEN, JF ;
MILLER, NC .
ANNALS OF BIOMEDICAL ENGINEERING, 1973, 1 (04) :455-467
[7]  
GUYTON AC, 1963, CIRCULATORY PHYSL CA, P163
[8]  
JABLKOWSKI J, 1980, T ISAO IFAC S, V2, P245
[9]   THEORETICAL CONSIDERATIONS REGARDING THE OPTIMIZATION OF CARDIAC ASSISTANCE BY INTRAAORTIC BALLOON PUMPING [J].
JARON, D ;
MOORE, TW ;
HE, P .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (03) :177-186
[10]  
KATZ AM, 1977, PHYSIOLOGY HEART, P397