EXCELLENT DURABILITY OF THE HANCOCK PORCINE BIOPROSTHESIS IN THE TRICUSPID POSITION

被引:26
作者
KAWACHI, Y
TOMINAGA, R
HISAHARA, M
NAKASHIMA, A
YASUI, H
TOKUNAGA, K
机构
关键词
D O I
10.1016/S0022-5223(19)33884-X
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
From February 1975 to August 1981, 23 consecutive patients underwent tricuspid valve replacement, which was either isolated (six patients) or combined with the replacement of other valves (17) by means of a standard, glutaraldehyde-preserved Hancock porcine bioprostheses. Patients' ages ranged from 9 to 53 (mean 36.2) years. The follow-up period ranged from 0.2 to 16.5 years (mean 9.1) and was complete in 100 % of all cases. Structural valve failure of the tricuspid Hancock valve was noticed in two patients, a 9-year-old boy and a 13-year-old girl 3.4 and 16.5 years after implantation, respectively. The actuarial freedom rate from structural valve failure at 10 years was 94 +/- 6%. There were six tricuspid prosthesis-related events: structural valve failure in two and valve thrombosis, anticoagulant-related bleeding, prosthetic valve endocarditis, and periprosthetic leak in one each, respectively. The actuarial freedom from these events at 10 years was 78 +/- 10%. Five pairs of aortic/mitral-tricuspid Hancock valves were explanted simultaneously from the same patients after 8.1 to 13.9 (mean 11.4) years postoperatively. A gross examination showed no valve dysfunction in the explants from the tricuspid position, but degenerative changes with valve dysfunction in those from the mitral and aortic position were observed (none of five versus five of seven; p < 0.03). We concluded that the selection of a Hancock bioprosthesis in the tricuspid position is acceptable because of the low incidence of prosthesis-related complications and the excellent durability of more than 10 years.
引用
收藏
页码:1561 / 1566
页数:6
相关论文
共 17 条
  • [1] COMPARISON OF LATE (62 TO 140 MONTHS) DEGENERATIVE CHANGES IN SIMULTANEOUSLY IMPLANTED AND EXPLANTED PORCINE (HANCOCK) BIOPROSTHESES IN THE TRICUSPID AND MITRAL-VALVE POSITIONS IN 6 PATIENTS
    COHEN, SR
    SILVER, MA
    MCINTOSH, CL
    ROBERTS, WC
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (11) : 1599 - 1602
  • [2] PORCINE VALVE DURABILITY IN CHILDREN
    DUNN, JM
    [J]. ANNALS OF THORACIC SURGERY, 1981, 32 (04) : 357 - 368
  • [3] GUIDELINES FOR REPORTING MORBIDITY AND MORTALITY AFTER CARDIAC VALVULAR OPERATIONS
    EDMUNDS, LH
    CLARK, RE
    COHN, LH
    MILLER, C
    WEISEL, RD
    [J]. ANNALS OF THORACIC SURGERY, 1988, 46 (03) : 257 - 259
  • [4] VALVE-REPLACEMENT IN THE RIGHT SIDE OF THE HEART IN CHILDREN - LONG-TERM FOLLOW-UP
    FLEMING, WH
    SARAFIAN, LB
    MOULTON, AL
    ROBINSON, LA
    KUGLER, JD
    [J]. ANNALS OF THORACIC SURGERY, 1989, 48 (03) : 404 - 408
  • [5] INTRINSIC FAILURE OF HANCOCK MITRAL BIOPROSTHESES - 10-YEAR TO 15-YEAR EXPERIENCE
    FOSTER, AH
    GREENBERG, GJ
    UNDERHILL, DJ
    MCINTOSH, CL
    CLARK, RE
    [J]. ANNALS OF THORACIC SURGERY, 1987, 44 (06) : 568 - 577
  • [6] STATISTICAL CONSIDERATIONS IN ANALYSIS AND REPORTING OF TIME-RELATED EVENTS - APPLICATION TO ANALYSIS OF PROSTHETIC VALVE-RELATED THROMBOEMBOLISM AND PACEMAKER FAILURE
    GRUNKEMEIER, GL
    THOMAS, DR
    STARR, A
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1977, 39 (02) : 257 - 258
  • [7] GUERRA F, 1990, J THORAC CARDIOV SUR, V99, P838
  • [8] JAMIESON WRE, 1990, J THORAC CARDIOV SUR, V99, P543
  • [9] KADO H, 1981, J JPN ASS THOR SURG, V29, P131
  • [10] COMPARATIVE-STUDY BETWEEN ST-JUDE MADICAL AND BIOPROSTHETIC VALVES IN THE RIGHT SIDE OF THE HEART
    KAWACHI, Y
    MASUDA, M
    TOMINAGA, R
    TOKUNAGA, K
    [J]. JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1991, 55 (06): : 553 - 562