SOFT-X-RAY BEAMLINE BL7A AT THE UVSOR

被引:35
作者
MURATA, T
MATSUKAWA, T
NAOE, S
HORIGOME, T
MATSUDO, O
WATANABE, M
机构
[1] NARUTO UNIV EDUC,DEPT PHYS,NARUTO 772,JAPAN
[2] KANAZAWA UNIV,COLL LIBERAL ARTS,DEPT PHYS,KANAZAWA,ISHIKAWA 920,JAPAN
[3] INST MOLEC SCI,OKAZAKI,AICHI 444,JAPAN
关键词
D O I
10.1063/1.1143054
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A vacuum-compatible double-crystal monochromator with constant exit-beam height has been installed and operated for several years at beamline BL7A at the UVSOR facility at the Institute for Molecular Science, Okazaki, Japan, The beamline is used both for the photons from a normal bending-magnet section and those from a 4 T superconducting three-pole horizontal wiggler. Various pairs of monochromator crystals are being used with sufficient intensity and signal to noise ratio of the output signal. Basic structure of the beamline, the mechanism of the monochromator, and some typical spectra of materials with absorption edges between 850 eV and 4 keV are reported.
引用
收藏
页码:1309 / 1312
页数:4
相关论文
共 5 条
[1]   A UHV COMPATIBLE 2-CRYSTAL MONOCHROMATOR FOR SYNCHROTRON RADIATION [J].
COWAN, PL ;
HASTINGS, JB ;
JACH, T ;
KIRKLAND, JP .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 208 (1-3) :349-353
[2]   X-RAY MONOCHROMATOR SYSTEM FOR USE WITH SYNCHROTRON RADIATION SOURCES [J].
GOLOVCHENKO, JA ;
LEVESQUE, RA ;
COWAN, PL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1981, 52 (04) :509-516
[3]  
Henke B. L., 1982, Atomic Data and Nuclear Data Tables, V27, P1, DOI 10.1016/0092-640X(82)90002-X
[4]  
MURATA T, 1986, J PHYS C SOLID STATE, V8, P135
[5]  
NAGAKURA I, 1964, SCI REP TOHOKU U, V48, P37