RADIOSURGERY WITH PHOTON BEAMS - PHYSICAL ASPECTS AND ADEQUACY OF LINEAR ACCELERATORS

被引:32
作者
PODGORSAK, EB [1 ]
PIKE, GB [1 ]
PLA, M [1 ]
OLIVIER, A [1 ]
SOUHAMI, L [1 ]
机构
[1] MCGILL UNIV,DEPT NEUROSURG,MONTREAL H3G 1A4,QUEBEC,CANADA
关键词
3-D dose calculations; Linear accelerators; Radiosurgery; Small field radiotherapy; Stereotaxy;
D O I
10.1016/0167-8140(90)90008-K
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The question of the adequacy of isocentric linear accelerators (linacs) for use in radiosurgery is addressed. The general physical requirements for radiosurgery, mainly a high spatial and numerical accuracy of dose delivery, reasonable treatment time, and low skin and leakage dose as well as cost considerations are examined. Various linac-based procedures are analyzed in view of their ability to meet these requirements and are contrasted with the clinically proven system of the Gamma unit. It is shown that the linac-based multiple converging arcs techniques and the dynamic rotation meet the stringent physical requirements on dose delivery and are thus viable alternatives to radiosurgery with the commercially available and dedicated Gamma unit. © 1990.
引用
收藏
页码:349 / 358
页数:10
相关论文
共 15 条
[1]  
Betti OO, 1984, ACTA NEUROCHIRURGI S, V33, P385
[2]   EXTERNAL STEREOTACTIC IRRADIATION BY LINEAR-ACCELERATOR [J].
COLOMBO, F ;
BENEDETTI, A ;
POZZA, F ;
AVANZO, RC ;
MARCHETTI, C ;
CHIEREGO, G ;
ZANARDO, A .
NEUROSURGERY, 1985, 16 (02) :154-160
[3]   CEREBRAL RADIATION SURGERY USING MOVING FIELD IRRADIATION AT A LINEAR-ACCELERATOR FACILITY [J].
HARTMANN, GH ;
SCHLEGEL, W ;
STURM, V ;
KOBER, B ;
PASTYR, O ;
LORENZ, WJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (06) :1185-1192
[4]   STEREOTAXIC RADIOTHERAPY TECHNIQUE FOR SMALL INTRACRANIAL LESIONS [J].
HOUDEK, PV ;
FAYOS, JV ;
VANBUREN, JM ;
GINSBERG, MS .
MEDICAL PHYSICS, 1985, 12 (04) :469-472
[5]   BRAGG-PEAK PROTON-BEAM THERAPY FOR ARTERIOVENOUS-MALFORMATIONS OF THE BRAIN [J].
KJELLBERG, RN ;
HANAMURA, T ;
DAVIS, KR ;
LYONS, SL ;
ADAMS, RD .
NEW ENGLAND JOURNAL OF MEDICINE, 1983, 309 (05) :269-274
[6]   A SYSTEM FOR STEREOTACTIC RADIOSURGERY WITH A LINEAR-ACCELERATOR [J].
LUTZ, W ;
WINSTON, KR ;
MALEKI, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (02) :373-381
[7]   STORAGE AN MANIPULATION OF RADIATION-DOSE DATA IN A SMALL DIGITAL COMPUTER [J].
MILAN, J ;
BENTLEY, RE .
BRITISH JOURNAL OF RADIOLOGY, 1974, 47 (554) :115-121
[8]  
OLIVIER A, 1986, APPL NEUROPHYSIOL, V48, P94
[9]   INTEGRATED STEREOTAXIC IMAGING WITH CT, MR IMAGING, AND DIGITAL SUBTRACTION ANGIOGRAPHY [J].
PETERS, TM ;
CLARK, JA ;
OLIVIER, A ;
MARCHAND, EP ;
MAWKO, G ;
DIEUMEGARDE, M ;
MURESAN, LV ;
ETHIER, R .
RADIOLOGY, 1986, 161 (03) :821-826
[10]   DOSE DISTRIBUTIONS IN DYNAMIC STEREOTAXIC RADIOSURGERY [J].
PIKE, B ;
PODGORSAK, EB ;
PETERS, TM ;
PLA, C .
MEDICAL PHYSICS, 1987, 14 (05) :780-789