A COMPUTER-CONTROLLED VERSATILE PULSE DUPLICATOR FOR PRECISION TESTING OF ARTIFICIAL-HEART VALVES

被引:21
作者
SCHICHL, K [1 ]
AFFELD, K [1 ]
机构
[1] FREE UNIV BERLIN,KLINIKUM RUDOLF VIRCHOW,W-1000 BERLIN,GERMANY
关键词
ARTIFICIAL HEART VALVES; PULSE DUPLICATOR; COMPUTER CONTROLLED; PRECISION TESTING; FLOW CURVE; VOLUMETRIC MEASUREMENT; PRESSURE DIFFERENCE CURVE;
D O I
10.1177/039139889301601009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Numerous devices and mock circulations have been described for the measurement of pressure loss, closure time, closing and leakage volumes and energy loss in artificial heart valves. However, all the devices have been troubled with difficulties in generating and assessing the precise flow through the valve, and problems in defining the arterial load, i.e. the artificial aorta. The new test device follows a radically different approach: a computer controlled piston forces the fluid through the test valve only - with no afterload. During systole, outflow follows a physiological curve which is identical for all types of heart valves of a given size. During diastole a mathematically defined physiological pressure difference curve is followed. Consequently, the measurements are independent of the individual machine, the lab where testing takes place, the scientist who executes the test, the time when measurements are taken and all other external influences.
引用
收藏
页码:722 / 728
页数:7
相关论文
共 8 条
[1]  
AFFELD K, 1985, Patent No. 3325179
[2]  
AFFELD K, 1986, LIFE SUPP SYST, V2, P142
[3]   PROSTHETIC HEART-VALVE EVALUATION INVITRO - CRITICAL ASPECTS OF DATA COMPARABILITY [J].
BARBARO, V ;
BARTOLINI, P ;
DANIELE, C ;
GRIGIONI, M ;
PALOMBO, A .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1991, 14 (06) :343-349
[4]   INVITRO EVALUATION OF PROSTHETIC HEART-VALVES - TOWARDS COMPARABLE TESTING [J].
BARBARO, V ;
DANIELE, C ;
GRIGIONI, M ;
PALOMBO, A ;
SARGENTINI, A .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1992, 16 (01) :10-14
[5]  
REUL H, 1983, ADV CARDIOVASCULAR P, V5, P17
[6]  
SCOTTEN LN, 1983, 18TH P AAMI ANN M
[7]   PROSTHETIC HEART-VALVES - WAVE-FORM COMPARISONS AND AVERAGE VALUE DISPOSITION FOR PULSATILE FLOW INVITRO [J].
SHE, JW ;
LI, MY ;
WANG, KQ ;
YU, YH .
JOURNAL OF BIOMEDICAL ENGINEERING, 1990, 12 (05) :375-378
[8]  
1989, ISO5840