HISTOLOGY OF POROUS-COATED ACETABULAR COMPONENTS - 25 CEMENTLESS CUPS RETRIEVED AFTER ARTHROPLASTY

被引:65
作者
SUMNER, DR [1 ]
JASTY, M [1 ]
JACOBS, JJ [1 ]
URBAN, RM [1 ]
BRAGDON, CR [1 ]
HARRIS, WH [1 ]
GALANTE, JO [1 ]
机构
[1] MASSACHUSETTS GEN HOSP, BOSTON, MA 02114 USA
来源
ACTA ORTHOPAEDICA SCANDINAVICA | 1993年 / 64卷 / 06期
关键词
D O I
10.3109/17453679308994584
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
25 hemispherical acetabular components retrieved from patients who had a total hip arthroplasty (THA) were studied histologically. All of the components had titanium fiber metal porous coatings and were implanted with adjuvant screw fixation, 18 in primary THA and 7 in revision THA. None of the cups had been removed because of infection or failure of fixation. The components had been in place an average of 30 (1-140) weeks. 18 cups had bone ingrowth into the porous coating. Up to one third of the available void space within the porous coating was occupied by bone, while the maximal proportion of the area of the bone-implant interface with bone ingrowth was over 80 percent. Bone ingrowth was more often observed at the dome and in the vicinity of sites of screw fixation than elsewhere. Peripherally, bone ingrowth was less consistent, and in isolated areas fibrocartilage and mineralized fibrocartilage were occasionally observed. The longer-term cases had more bone ingrowth than the shorter-term cases. Apparent metallic debris was observed within histiocytes adjacent to screw holes, and the amount of this material increased with duration of implantation.
引用
收藏
页码:619 / 626
页数:8
相关论文
共 20 条
[1]  
[Anonymous], 1992, SOCIOLOGICA, V7
[2]  
COLLIER JP, 1988, CLIN ORTHOP RELAT R, V235, P173
[3]  
Cook Stephen D, 1991, J LONG-TERM EFF MED, V1, P11
[4]  
DELEE JG, 1976, CLIN ORTHOP RELAT R, P20
[5]  
DICARLO E F, 1992, Clinical Materials, V9, P235, DOI 10.1016/0267-6605(92)90104-2
[6]  
GLANT TT, 1993, J BONE MINER RES, V8, P1071
[7]  
Harris W H, 1985, Hip, P214
[8]  
JACOBS JJ, 1992, AM SOC TEST MATER, V1144, P52, DOI 10.1520/STP14851S
[9]   INGROWTH AND FORMATION OF BONE IN DEFECTS IN AN UNCEMENTED FIBER-METAL TOTAL HIP-REPLACEMENT MODEL IN DOGS [J].
KANG, JD ;
MCKERNAN, DJ ;
KRUGER, M ;
MUTSCHLER, T ;
THOMPSON, WH ;
RUBASH, HE .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (01) :93-105
[10]   ENDOSTEAL EROSION IN ASSOCIATION WITH STABLE UNCEMENTED FEMORAL COMPONENTS [J].
MALONEY, WJ ;
JASTY, M ;
HARRIS, WH ;
GALANTE, JO ;
CALLAGHAN, JJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (07) :1025-1034