THE DETERMINATION OF THE THERMODYNAMIC TEMPERATURES OF THERMOMETRY FIXED-POINTS IN THE RANGE 660-DEGREES-C TO 1064-DEGREES-C

被引:19
作者
JONES, TP
TAPPING, J
机构
[1] CSIRO, Lindfield, Aust, CSIRO, Lindfield, Aust
关键词
COPPER SILVER ALLOYS - Thermodynamic Properties - GOLD AND ALLOYS - Thermodynamic Properties - SILVER AND ALLOYS - Thermodynamic Properties - TEMPERATURE MEASUREMENT;
D O I
10.1088/0026-1394/25/1/007
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The thermodynamic temperatures of silver, gold and copper-silver eutectic fixed points were measured with a photoelectric pyrometer relative to an aluminum point of 660. 302 degree C. The values obtained were (961. 744 plus or minus 0. 055) degree C, (1064. 145 plus or minus 0. 060) degree C and (779. 583 plus or minus 0. 060) degree C respectively. Although the final uncertaint of the copper-silver eutectic was similar to that of the other two, the effort necessary to obtain it was much greater and it is not recommended as a major reference point for use in photoelectric pyrometry.
引用
收藏
页码:41 / 47
页数:7
相关论文
共 21 条
[1]  
ANDREWS JW, 1984, BIPM COM CONS THERMO, V15
[2]  
Bedford R. E., 1982, TEMPERATURE ITS MEAS, V5, P361
[3]   THE REPORTING OF THERMODYNAMIC TEMPERATURE INTERVALS MEASURED WITH AN OPTICAL-PYROMETER [J].
BEDFORD, RE .
METROLOGIA, 1982, 18 (03) :169-170
[4]   RADIOMETRIC DETERMINATION OF THERMODYNAMIC TEMPERATURES BETWEEN 904 AND 1338 K [J].
BONHOURE, J .
METROLOGIA, 1975, 11 (04) :141-150
[5]  
COATES PB, 1982, TEMP ITS MEAS, V5, P109
[6]   RADIOMETRIC DETERMINATION OF THERMODYNAMIC TEMPERATURE INTERVAL BETWEEN SILVER AND GOLD FREEZING POINTS [J].
COSLOVI, L ;
ROSSO, A ;
RUFFINO, G .
METROLOGIA, 1975, 11 (02) :85-87
[7]   DEVIATION OF INTERNATIONAL PRACTICAL TEMPERATURES FROM THERMODYNAMIC TEMPERATURES IN TEMPERATURE-RANGE FROM 273.16K TO 730 K [J].
GUILDNER, LA ;
EDSINGER, RE .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS SECTION A-PHYSICS AND CHEMISTRY, 1976, 80 (5-6) :703-738
[8]  
HUDSON RP, 1987, BIPM COM CONS THERMO, V16
[9]   REALIZATION OF IPTS68 ABOVE 1064.43 DEGREE C USING NSL PHOTOELECTRIC PYROMETER [J].
JONES, TP ;
TAPPING, J .
METROLOGIA, 1972, 8 (01) :4-+
[10]   A PRECISION PHOTOELECTRIC PYROMETER FOR THE REALIZATION OF THE IPTS-68 ABOVE 1064.43-DEGREES-C [J].
JONES, TP ;
TAPPING, J .
METROLOGIA, 1982, 18 (01) :23-31