ONLINE STANDARD ADDITIONS WITH DIRECT INJECTION NEBULIZATION FOR INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY

被引:56
作者
WIEDERIN, DR
SMYCZEK, RE
HOUK, RS
机构
[1] IOWA STATE UNIV SCI & TECHNOL,US DOE,AMES LAB,AMES,IA 50011
[2] IOWA STATE UNIV SCI & TECHNOL,DEPT CHEM,AMES,IA 50011
关键词
D O I
10.1021/ac00015a024
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow injection method for on-line standard additions is described. A 500-mu-L sample solution is injected into a flowing stream via a metal-free flow injection valve. Standard solutions (50-mu-L) are sequentially introduced through a second valve. The two streams mix in a PEEK tee before they are nebulized into the ICP. The advantages of the method of standard additions for matrix correction are retained, but the time-consuming solution preparation steps for conventional standard additions are eliminated by using flow injection techniques and a direct injection nebulizer. Nine toxic elements in undiluted urine were determined in less than 5 min. Possible effects of sample viscosity on accuracy are investigated. Evidence for a change in the background when seawater is nebulized is also presented.
引用
收藏
页码:1626 / 1631
页数:6
相关论文
共 28 条