DETECTION AND DETERMINATION LIMITS FOR THERMOLUMINESCENCE DOSIMETRY

被引:40
作者
HIRNING, CR
机构
[1] Health and Safety Division, Ontario Hydro, Pickering, ON, LlW 3C8
来源
HEALTH PHYSICS | 1992年 / 62卷 / 03期
关键词
THERMOLUMINESCENCE DOSIMETRY; DOSIMETRY; PERSONNEL; NUCLEAR WORKERS;
D O I
10.1097/00004032-199203000-00002
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The statistical theory of detection and determination limits is applied to thermoluminescence dosimetry (TLD), and equations are derived for calculating the limits. Three different levels of testing a TLD system are described, and a test procedure is given for each level. An example from a working system is presented, and the dependence of detection and determination limits on background radiation exposure is shown. The results show the importance of clearly defining the terms used to describe the low-dose performance of a TLD system.
引用
收藏
页码:223 / 227
页数:5