SPIN-LATTICE RELAXATION-TIME MEASUREMENT BY MEANS OF A TURBOFLASH TECHNIQUE

被引:111
作者
BLUML, S
SCHAD, LR
STEPANOW, B
LORENZ, WJ
机构
[1] Department of Radiological Diagnostic and Therapy, German Cancer Research Center (Dkfz), Heidelberg
关键词
MAGNETIC RESONANCE IMAGING; RELAXATION MECHANISM; SPIN-LATTICE RELAXATION (T(1)); TISSUE CHARACTERIZATION;
D O I
10.1002/mrm.1910300304
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rapid measurements of in vivo proton spin-lattice relaxation times (T1) in human tissues were performed by magnetic resonance imaging in a 1.5 T whole-body super-conducting MR scanner. The measurements employ serial TurboFLASH imaging (Snapshot-FLASH) with scan times for a single experiment below 4 s. Using centric phase encoding order, an appropriate fitting of the T1-parameter from images with minimum motion artifacts is possible. Comparative T1-determination with a multipoint inversion recovery and spin-echo technique was performed on phantoms containing Gd-DTPA solutions with different T1-values. We found a maximum deviation of 3.3% for T1 < 11 00 ms and of 12.1% for 11 00 ms < T1 < 1700 ms from the results obtained by the IR technique. In vivo measurements of T1-relaxation times were performed in white and grey matter, cerebrospinal fluid, kidney, liver, spleen, muscle, and bone marrow, and yielded values that are in good agreement with reported data.
引用
收藏
页码:289 / 295
页数:7
相关论文
共 17 条
  • [1] Abragam A., 1961, PRINCIPLES NUCLEAR M, P264
  • [2] BOTTOMLEY PA, 1984, MED PHYS, V11, P425, DOI 10.1118/1.595535
  • [3] FAST AND PRECISE T1 IMAGING USING A TOMROP SEQUENCE
    BRIX, G
    SCHAD, LR
    DEIMLING, M
    LORENZ, WJ
    [J]. MAGNETIC RESONANCE IMAGING, 1990, 8 (04) : 351 - 356
  • [4] CHIEN D, 1991, Magnetic Resonance Quarterly, V7, P31
  • [5] A COMPARISON OF ONE-SHOT AND RECOVERY METHODS IN T1-IMAGING
    CRAWLEY, AP
    HENKELMAN, RM
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1988, 7 (01) : 23 - 34
  • [6] DEICHMANN R, 1991, BOOK ABSTRACTS SOC M, P699
  • [7] FRAHM KE, 1987, MAGN RESON IMAGING, V5, P201
  • [8] GRAUMANN R, 1986, 5TH P ANN M SOC MAG, P922
  • [9] SNAPSHOT FLASH MRI - APPLICATIONS TO T1, T2, AND CHEMICAL-SHIFT IMAGING
    HAASE, A
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 13 (01) : 77 - 89
  • [10] FLASH IMAGING - RAPID NMR IMAGING USING LOW FLIP-ANGLE PULSES
    HAASE, A
    FRAHM, J
    MATTHAEI, D
    HANICKE, W
    MERBOLDT, KD
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1986, 67 (02) : 258 - 266