NUCLEAR SPECTROSCOPY WITH SI PIN DIODE DETECTORS AT ROOM-TEMPERATURE

被引:27
作者
AHMAD, I
BETTS, RR
HAPP, T
HENDERSON, DJ
WOLFS, FLH
WUOSMAA, AH
机构
[1] Physics Division, Argonne National Laboratory, Argonne, Il 60439
关键词
D O I
10.1016/0168-9002(90)90776-3
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The characteristics of PIN diodes have been determined. These diodes have lower leakage currents and noise than other types of Si radiation detectors. The energy resolutions (FWHM) of a 1 cm2 x 0.5 mm PIN diode measured with a pulser, 122.0 keV gamma-rays, 193 keV electrons and 5.5 MeV alpha-particles were 2.6, 2.8, 2.9 and 11.0 keV, respectively. For a 6 mm x 6 mm x 0.2 mm PIN diode, the resolutions (FWHM) for a pulser, 60 keV gamma-rays, 193 keV electrons and 5.5-Mev alpha-particles were 2.1, 2.2, 2.4 and 10.8 keV, respectively.
引用
收藏
页码:201 / 204
页数:4
相关论文
共 6 条
[1]   SIMPLE COOLED SI(LI) ELECTRON SPECTROMETER [J].
AHMAD, I ;
WAGNER, F .
NUCLEAR INSTRUMENTS & METHODS, 1974, 116 (03) :465-469
[3]  
KEMMER J, 1984, NUCL INSTRUM METH A, V226, P89, DOI 10.1016/0168-9002(84)90173-6
[4]  
KNOLL GF, 1989, RAD DETECTION MEASUR, P400
[5]   LOW-ENERGY NUCLEAR RADIATION DETECTION WITH A SILICON PHOTODIODE [J].
YAMAMOTO, H ;
HATAKEYAMA, S ;
NORIMURA, T ;
TSUCHIYA, T .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 281 (01) :128-132
[6]   LARGE AREA MOSAIC SOLID-STATE DETECTORS FOR LOW YIELD FISSION STUDY [J].
YAMAMOTO, H ;
MORI, Y ;
WAKUTA, Y ;
SONODA, M .
NUCLEAR INSTRUMENTS & METHODS, 1976, 134 (01) :119-124