CHEMILUMINESCENT FLOW SENSOR FOR THE DETERMINATION OF PARAOXON AND ALDICARB PESTICIDES

被引:59
作者
RODA, A
RAUCH, P
FERRI, E
GIROTTI, S
GHINI, S
CARREA, G
BOVARA, R
机构
[1] UNIV BOLOGNA,INST CHEM SCI,I-40127 BOLOGNA,ITALY
[2] UNIV BOLOGNA,DEPT PHARMACEUT SCI,I-40126 BOLOGNA,ITALY
[3] PRAGUE INST CHEM TECHNOL,DEPT BIOCHEM & MICROBIOL,CR-16628 PRAGUE,CZECH REPUBLIC
[4] CNR,INST HORMONE CHEM,I-20131 MILAN,ITALY
关键词
CHEMILUMINESCENCE; FLOW SYSTEM; CARBAMATES; ORGANOPHOSPHORUS COMPOUNDS; PESTICIDES;
D O I
10.1016/0003-2670(94)85043-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A chemiluminescence-based flow method for the determination of some organophorus and carbamate pesticides based on inhibition of acetylcholinesterase was developed. Acetylcholinesterase in solution or immobilized on methacrylate beads (Eupergit C) was coupled to choline oxidase and peroxidase immobilized on Eupergit C. In this system choline formed by acetylcholinesterase was oxidized by choline oxidase and the H2O2 produced was determined via the luminol/peroxidase luminescent reaction. The detection limits (3 sigma) for Paraoxon and Aldicarb were 0.75 mu g l(-1) and 4 mu g l(-1), respectively, when soluble acetylcholinesterase was used under the following optimized experimental conditions: 56 mu M luminol in working solution, sample volume 60 mu l, flow-rate 0.3 ml min(-1) and 60 min incubation time. The flow sensor device using all the three enzymes in the immobilized form had a higher detection limit of 125 mu g l(-1) for Paraoxon. The mid-range relative standard deviation (n = 10) using 1 mM standard substrate solution was 3.7%. The recovery from contaminated samples (soil, vegetables) varied from 81 to 108%. The results obtained by the developed methods were in good agreement with those obtained by a commonly used colorimetric test.
引用
收藏
页码:35 / 42
页数:8
相关论文
共 25 条
[1]  
Bergmeyer H.U., 1984, METHODDS ENZYMATIC A, P52
[2]  
BERGMEYER HU, 1983, METHOD ENZYMATIC ANA, V2, P128
[3]   ANTICHOLINESTERASE ACTIVITY MEASUREMENT BY A CHOLINE BIOSENSOR - APPLICATION IN WATER ANALYSIS [J].
BERNABEI, M ;
CHIAVARINI, S ;
CREMISINI, C ;
PALLESCHI, G .
BIOSENSORS & BIOELECTRONICS, 1993, 8 (05) :265-271
[4]   DETERMINATION OF ORGANOPHOSPHORUS AND CARBAMIC PESTICIDES WITH A CHOLINE AND ACETYLCHOLINE ELECTROCHEMICAL BIOSENSOR [J].
BERNABEI, M ;
CREMISINI, C ;
MASCINI, M ;
PALLESCHI, G .
ANALYTICAL LETTERS, 1991, 24 (08) :1317-1331
[5]   SENSITIVE AND RAPID DETECTION OF PARAOXON BY THIN-LAYER CHROMATOGRAPHY AND STRIPS USING ENZYME-INHIBITION AND ELLMAN METHOD [J].
BREUER, H .
JOURNAL OF CHROMATOGRAPHY, 1982, 243 (01) :183-187
[6]   BUTYRYLCHOLINE ENZYME SENSOR FOR DETERMINING ORGANOPHOSPHORUS INHIBITORS [J].
CAMPANELLA, L ;
ACHILLI, M ;
SAMMARTINO, MP ;
TOMASSETTI, M .
BIOELECTROCHEMISTRY AND BIOENERGETICS, 1991, 26 (02) :237-249
[7]  
Coats J. R, 1982, INSECTICIDE MODE ACT
[8]  
Colliss Janet S., 1993, Analytical Proceedings, V30, P74
[9]  
Corbett J. R., 1974, BIOCH MODE ACTION PE
[10]  
DAHM PA, 1970, BIOCH TOXICOLOGY INS