AUTOMATIC, NONINVASIVE INTRADIALYTIC CLEARANCE MEASUREMENT

被引:111
作者
POLASCHEGG, HD [1 ]
机构
[1] FRESENIUS AG,DIV RES & DEV MED SYST,HOMBURG,GERMANY
关键词
HEMODIALYSIS; QUALITY CONTROL; EFFICACY; CLEARANCE; DIALYSANCE;
D O I
10.1177/039139889301600404
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The feasibility of non invasive, automatic intradialytic measurement of effective clearance is demonstrated by a mathematical derivation. In a practical setting the conductivity difference between dialysate inlet and outlet of the dialyzer is measured for two different dialysate inlet concentrations. This allows for the calculation of the electrolyte dialysance, which is equal to the effective urea clearance, provided dialysate flow, blood flow and blood electrolyte composition remain unchanged during the measuring interval
引用
收藏
页码:185 / 191
页数:7
相关论文
共 10 条
[1]  
ALBERS JR, 1965, ASAIO T, V11, P161
[2]  
BOAG J, 1987, CONT DIAL NEPHROL, V8, P11
[3]   A MECHANISTIC ANALYSIS OF THE NATIONAL COOPERATIVE DIALYSIS STUDY (NCDS) [J].
GOTCH, FA ;
SARGENT, JA .
KIDNEY INTERNATIONAL, 1985, 28 (03) :526-534
[4]  
GOTCH FA, 1984, 1 INT S SINGL NEEDL, P47
[5]  
POLASCHEGG HD, 1982, Patent No. 97366
[6]  
POLASCHEGG HD, 1991, Patent No. 91108005
[7]  
POLASCHEGG HD, 1989, Patent No. 3938662
[8]  
Sargent JA, 1983, REPLACEMENT RENAL FU, P53
[9]  
Schneditz D, 1992, ASAIO J, V38, pM194, DOI 10.1097/00002480-199207000-00018
[10]  
STEIL H, IN PRESS