REDUCTION OF ACOUSTIC NOISE IN ICP EMISSION BY THE ADDITION OF A CHIMNEY

被引:15
作者
EASLEY, SF [1 ]
MONNIG, CA [1 ]
HIEFTJE, GM [1 ]
机构
[1] INDIANA UNIV,DEPT CHEM,BLOOMINGTON,IN 47405
关键词
NOISE-REDUCTION; INDUCTIVELY COUPLED PLASMA; ATOMIC EMISSION SPECTROSCOPY;
D O I
10.1366/0003702914335869
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Noise-amplitude spectra were measured for inductively coupled plasma (ICP) atomic emission spectrometry. These spectra indicate that a significant reduction of noise in the emission signal can be obtained by positioning a chimney over the ICP discharge. In this fashion, noise was reduced in the 200-500 Hz region of the frequency spectrum. The chimney had little effect on flicker (1/f) noise.
引用
收藏
页码:1368 / 1371
页数:4
相关论文