SOURCE DISTRIBUTION DEPENDENT SCATTER CORRECTION FOR PVI

被引:12
作者
BARNEY, JS [1 ]
HARROP, R [1 ]
DYKSTRA, CJ [1 ]
机构
[1] TRIUMF,VANCOUVER V6T 2A3,BC,CANADA
基金
加拿大自然科学与工程研究理事会;
关键词
D O I
10.1109/23.256701
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Source distribution dependent scatter correction methods which incorporate different amounts of information about. the source position and material distribution have been developed and tested. The techniques use image to projection integral transformation incorporating varying degrees of information on the distribution of scattering material, or convolution subtraction methods, with some information about the scattering material included in one of the convolution methods. To test the techniques, we apply them to data generated by Monte Carlo simulations which use geometric shapes or a voxelized density map to model the scattering material. Source position and material distribution have been found to have some effect on scatter correction. An image to projection method which incorporates a density map produces accurate scatter correction but is computationally expensive. Simpler methods, both image to projection and convolution, can also provide effective scatter correction. 1
引用
收藏
页码:1001 / 1007
页数:7
相关论文
共 15 条
[1]  
BARNEY J, 1993, THESIS S FRASER U
[2]   OBJECT SHAPE DEPENDENT SCATTER SIMULATIONS FOR PET [J].
BARNEY, JS ;
ROGERS, JG ;
HARROP, R ;
HOVERATH, H .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1991, 38 (02) :719-725
[3]   CORRECTION FOR SCATTERED RADIATION IN A RING DETECTOR POSITRON CAMERA BY INTEGRAL TRANSFORMATION OF THE PROJECTIONS [J].
BERGSTROM, M ;
ERIKSSON, L ;
BOHM, C ;
BLOMQVIST, G ;
LITTON, J .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1983, 7 (01) :42-50
[4]  
GROOTOONK S, 1991, 1991 IEEE NUCL SCI S, P1569
[5]  
Kak A. C., 1999, PRINCIPLES COMPUTERI
[6]  
KARP JS, 1990, J NUCL MED, V31, P617
[7]   COMPTON-SCATTERING IN A LARGE-APERTURE POSITRON IMAGING-SYSTEM [J].
MCKEE, BTA ;
HOGAN, MJ ;
HOWSE, DCN .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (03) :198-202
[8]   ASSESSMENT OF RESPONSE FUNCTION IN 2 PET SCANNERS WITH AND WITHOUT INTERPLANE SEPTA [J].
MICHEL, C ;
BOL, A ;
SPINKS, T ;
TOWNSEND, D ;
BAILEY, D ;
GROOTOONK, S ;
JONES, T .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) :240-248
[9]  
OLLINGER J M, 1991, Journal of Nuclear Medicine, V32, P996
[10]  
OLLINGER JM, 1992, 1992 IEEE NUCL SCI S