MULTI-MOTOR UNIT ACTION-POTENTIAL ANALYSIS (MMA)

被引:53
作者
NANDEDKAR, SD
BARKHAUS, PE
CHARLES, A
机构
[1] CTR ATOM SIMULAT & ANAL,HOPEWELL JCT,NY
[2] MED COLL WISCONSIN,DEPT NEUROL,MILWAUKEE,WI 53226
[3] MEDELEC LTD,SURREY,ENGLAND
关键词
MOTOR UNIT ACTION POTENTIALS; QUANTITATIVE EMG; INTERFERENCE PATTERN; DECOMPOSITION; COMPUTER-AIDED EMG;
D O I
10.1002/mus.880181012
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
We have developed an algorithm, called multi-motor unit action potential analysis (MMA), to aid quantification in routine needle EMG examination, In only 5-8 min, it was possible to extract, analyze, and validate 20 motor unit action potentials (MUAPs). In the biceps muscle of normal subjects, the MUAP measurements are compared with measurements using manual, triggered averaging, automatic decomposition, and other MMA algorithms described in the literature. The mean values of MUAP features by MMA fell between the extreme limits obtained from the amplitude-triggered method and the manual method described in the literature. In patients, the results are consistent with routine EMG and similar to those results described in the literature, Fast acquisition and analysis of MUAPs as well as interference pattern (IF) analysis make this technique useful to document EMG abnormalities in routine needle examination. (C) 1995 John Wiley & Sons, Inc.
引用
收藏
页码:1155 / 1166
页数:12
相关论文
共 18 条
[1]  
Andreassen S, 1987, Electroencephalogr Clin Neurophysiol Suppl, V39, P13
[2]   QUANTITATIVE EMG IN INFLAMMATORY MYOPATHY [J].
BARKHAUS, PE ;
NANDEDKAR, SD ;
SANDERS, DB .
MUSCLE & NERVE, 1990, 13 (03) :247-253
[3]   REFERENCE VALUES OF MOTOR UNIT ACTION-POTENTIALS OBTAINED WITH MULTI-MUAP ANALYSIS [J].
BISCHOFF, C ;
STALBERG, E ;
FALCK, B ;
EEGOLOFSSON, KE .
MUSCLE & NERVE, 1994, 17 (08) :842-851
[4]   THE DIAGNOSTIC YIELD OF QUANTIFIED ELECTROMYOGRAPHY AND QUANTIFIED MUSCLE BIOPSY IN NEUROMUSCULAR DISORDERS [J].
BUCHTHAL, F ;
KAMIENIECKA, Z .
MUSCLE & NERVE, 1982, 5 (04) :265-280
[5]  
DORFMAN L, 1989, COMPUTER AIDED ELECT, P189
[6]  
JACKSON LB, 1986, DIGITAL FILTERS SIGN
[7]  
McGill KC, 1989, COMPUTER AIDED ELECT, P91
[8]   SIMULATION OF CONCENTRIC NEEDLE EMG MOTOR UNIT ACTION-POTENTIALS [J].
NANDEDKAR, SD ;
SANDERS, DB ;
STALBERG, EV ;
ANDREASSEN, S .
MUSCLE & NERVE, 1988, 11 (02) :151-159
[9]   ANALYSIS OF AMPLITUDE AND AREA OF CONCENTRIC NEEDLE EMG MOTOR UNIT ACTION-POTENTIALS [J].
NANDEDKAR, SD ;
BARKHAUS, PE ;
SANDERS, DB ;
STALBERG, EV .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1988, 69 (06) :561-567
[10]  
Rosenfalck P., 1975, ELECTROMYOGRAPHY SEN