DETECTION OF GLUCOSE VIA ELECTROCHEMILUMINESCENCE IN A THIN-LAYER CELL WITH A PLANAR OPTICAL WAVE-GUIDE

被引:55
作者
KREMESKOTTER, J [1 ]
WILSON, R [1 ]
SCHIFFRIN, DJ [1 ]
LUFF, BJ [1 ]
WILKINSON, JS [1 ]
机构
[1] UNIV SOUTHAMPTON, OPTOELECTR RES CTR, SOUTHAMPTON SO17 1BJ, HANTS, ENGLAND
关键词
D O I
10.1088/0957-0233/6/9/012
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Light generated by luminol electrochemiluminescence is coupled into a simple planar optical waveguide and is collected with a photomultiplier tube and a photon counter unit. The waveguide is mounted to a thin-layer cell which is connected to a flow injection analysis system. The waveguide is coated with indium tin oxide and modified with covalently attached glucose oxidase. The range of detection for glucose 0-10 mM (correlation coefficient r = 0.9974), with a detection limit of 0.3 mM.