LEFT-VENTRICULAR VOLUME MEASUREMENT USING CARDIAC AXIS NUCLEAR-MAGNETIC-RESONANCE IMAGING - VALIDATION BY CALIBRATED VENTRICULAR ANGIOGRAPHY

被引:158
作者
CRANNEY, GB
LOTAN, CS
DEAN, L
BAXLEY, W
BOUCHARD, A
POHOST, GM
机构
[1] Division of Cardiovascular Disease, Department of Medicine, University of Alabama at Birmingham, Birmingham, AL
[2] Division of Cardiovascular Disease, University of Alabama at Birmingham, Birmingham
关键词
Cardiac volumes; Heart function; Left ventricle; Nuclear magnetic resonance;
D O I
10.1161/01.CIR.82.1.154
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Proton nuclear magnetic resonance (NMR) imaging has the potential to serially assess left ventricular (LV) volumes with optimal accuracy because it is a high-resolution, three-dimensional, noninvasive modality. Previous NMR studies to assess LV volumes have been suboptimal, as they have used either planes aligned with the axes of the body, which are compromised by partial volume effects, or spin-echo techniques that have been time-consuming to acquire and analyze. Accordingly, for LV volume measurement, we developed a gradient-echo (cine) NMR strategy that uses two orthogonal planes intersecting along the intrinsic long axis of the heart (two-chamber and four-chamber). This approach was validated against calibrated contrast biplane LV cineangiography (CATH) and also compared with a previously reported short-axis spin-echo NMR method. Twenty-one patients underwent CATH and NMR (long-axis, n=21; short-axis, n=14) within a 3-day interval. Although both long- and short-axis NMR LV volumes and ejection fractions correlated well with CATH (r>0.90, p<0.001 in all), end-diastolic volumes by both long-axis (161±85 ml) and short-axis (151±81 ml) NMR were systematically less than those by CATH (182±85 ml) (p<0.05). Consequently, ejection fractions by long-axis (48±17%) and short-axis (49±17%) NMR consistently underestimated those by CATH (54±16%, p<0.05). End-systolic volumes by long-axis (94±71 ml) and short-axis (87±72 ml) NMR were not significantly different from those by CATH (92±69 ml). Both NMR techniques had low intraobserver and interobserver variation (<11%); however, short-axis spin-echo NMR involved longer acquisition/reconstruction (35 versus 18 minutes) and analysis (25 versus 10 minutes) times. We conclude that both short-axis spin-echo and long-axis gradient-echo NMR approaches reliably estimate LV volumes. Currently, the long-axis strategy appears more practical for clinical use because the scan and analysis times are relatively short.
引用
收藏
页码:154 / 163
页数:10
相关论文
共 33 条
  • [1] END-SYSTOLIC VOLUME AS A PREDICTOR OF POSTOPERATIVE LEFT-VENTRICULAR PERFORMANCE IN VOLUME OVERLOAD FROM VALVULAR REGURGITATION
    BOROW, KM
    GREEN, LH
    MANN, T
    SLOSS, LJ
    BRAUNWALD, E
    COLLINS, JJ
    COHN, L
    GROSSMAN, W
    [J]. AMERICAN JOURNAL OF MEDICINE, 1980, 68 (05) : 655 - 663
  • [2] NORMAL CARDIAC DIMENSIONS BY MAGNETIC-RESONANCE IMAGING
    BYRD, BJ
    SCHILLER, NB
    BOTVINICK, EH
    HIGGINS, CB
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1985, 55 (11) : 1440 - 1442
  • [3] REAL-TIME MOVIE IMAGING FROM A SINGLE CARDIAC CYCLE BY NMR
    CHAPMAN, B
    TURNER, R
    ORDIDGE, RJ
    DOYLE, M
    CAWLEY, M
    COXON, R
    GLOVER, P
    MANSFIELD, P
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1987, 5 (03) : 246 - 254
  • [4] NON-GEOMETRIC DETERMINATION OF LEFT-VENTRICULAR VOLUMES FROM EQUILIBRIUM BLOOD POOL SCANS
    DEHMER, GJ
    LEWIS, SE
    HILLIS, LD
    TWIEG, D
    FALKOFF, M
    PARKEY, RW
    WILLERSON, JT
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1980, 45 (02) : 293 - 300
  • [5] DETERMINATION OF LEFT-VENTRICULAR VOLUMES AND EJECTION FRACTION BY NUCLEAR-MAGNETIC-RESONANCE IMAGING
    DILWORTH, LR
    AISEN, AM
    MANCINI, GBJ
    LANDE, I
    BUDA, AJ
    [J]. AMERICAN HEART JOURNAL, 1987, 113 (01) : 24 - 32
  • [7] THE USE OF BIPLANE ANGIOCARDIOGRAPHY FOR THE MEASUREMENT OF LEFT VENTRICULAR VOLUME IN MAN
    DODGE, HT
    SANDLER, H
    BALLEW, DW
    LORD, JD
    [J]. AMERICAN HEART JOURNAL, 1960, 60 (05) : 762 - 776
  • [8] CARDIAC-FUNCTION - EVALUATION WITH FAST-ECHO MR IMAGING
    EDELMAN, RR
    THOMPSON, R
    KANTOR, H
    BRADY, TJ
    LEAVITT, M
    DINSMORE, R
    [J]. RADIOLOGY, 1987, 162 (03) : 611 - 615
  • [9] COMPARISON OF MAGNETIC-RESONANCE IMAGING AND ECHOCARDIOGRAPHY IN DETERMINATION OF CARDIAC DIMENSIONS IN NORMAL SUBJECTS
    FRIEDMAN, BJ
    WATERS, J
    KWAN, OL
    DEMARIA, AN
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1985, 5 (06) : 1369 - 1376
  • [10] LONG-TERM VASODILATOR THERAPY OF CHRONIC AORTIC-INSUFFICIENCY - A RANDOMIZED DOUBLE-BLINDED, PLACEBO-CONTROLLED CLINICAL-TRIAL
    GREENBERG, B
    MASSIE, B
    BRISTOW, JD
    CHEITLIN, M
    SIEMIENCZUK, D
    TOPIC, N
    WILSON, RA
    SZLACHCIC, J
    THOMAS, D
    [J]. CIRCULATION, 1988, 78 (01) : 92 - 103