EARLY ASEPTIC LOOSENING OF UNCEMENTED FEMORAL COMPONENTS IN PRIMARY TOTAL HIP-REPLACEMENT - A REVIEW BASED ON THE NORWEGIAN ARTHROPLASTY REGISTER

被引:102
作者
HAVELIN, LI
ESPEHAUG, B
VOLLSET, SE
ENGESAETER, LB
机构
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1995年 / 77B卷 / 01期
关键词
D O I
10.1302/0301-620X.77B1.7822364
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The Norwegian Arthroplasty Register recorded 24 408 primary total hip replacements from 1987 to 1993; 2907 of them (13%) were performed with uncemented femoral components. We have compared the results of eight different designs, each used in more than 100 patients. Survivorship of the components was estimated by the Kaplan-Meier method using revision for aseptic loosening of the femoral component as the end-point. At 4.5 years, the estimated probability of revision for aseptic loosening for all implants was 4.5%, for the Bio-Fit stem 18.6% (n = 210) and for the Femora stem 13.6% (n = 173). The PM-Prosthesis and the Harris/Galante stem prostheses needed revision in 5.6% and 3.6%, respectively. The clockwise threaded stem of the Femora implant needed revision in 20% of right hips, but in only 4% of left hips. The short-term results of the four best uncemented femoral components (Corail, LMT, Profile and Zweimuller) were similar to those for cemented stems, with revision for loosening in less than 1% at 4.5 years. The importance of the control of innovative designs and the registration of early results is discussed.
引用
收藏
页码:11 / 17
页数:7
相关论文
共 19 条
[1]
AHNFELT L, 1990, ACTA ORTHOP SCAND, V61, P1
[2]
WHAT PRICE PROGRESS - FAILED INNOVATIONS OF THE KNEE PROSTHESIS [J].
BAUER, GCH .
ACTA ORTHOPAEDICA SCANDINAVICA, 1992, 63 (03) :245-246
[3]
COX DR, 1972, J R STAT SOC B, V34, P187
[4]
DIXON WJ, 1990, BMDP STATISTICAL SOF, V1
[5]
RESULTS OF 203 TOTAL HIP REPLACEMENTS USING A SMOOTH, CEMENTLESS FEMORAL COMPONENT [J].
DUPARC, J ;
MASSIN, P .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (02) :251-256
[6]
FARO LMC, 1992, ACTA ORTHOP SC S250, V63
[7]
THE NORWEGIAN-ARTHROPLASTY-REGISTER - A SURVEY OF 17,444 HIP REPLACEMENTS 1987-1990 [J].
HAVELIN, LI ;
ESPEHAUG, B ;
VOLLSET, SE ;
ENGESAETER, LB ;
LANGELAND, N .
ACTA ORTHOPAEDICA SCANDINAVICA, 1993, 64 (03) :245-251
[8]
EARLY FAILURES AMONG 14,009 CEMENTED AND 1,326 UNCEMENTED PROSTHESES FOR PRIMARY COXARTHROSIS - THE NORWEGIAN ARTHROPLASTY REGISTER, 1987-1992 [J].
HAVELIN, LI ;
ESPEHAUG, B ;
VOLLSET, SE ;
ENGESAETER, LB .
ACTA ORTHOPAEDICA SCANDINAVICA, 1994, 65 (01) :1-6
[9]
HAVELIN LI, 1991, P NORWEGIAN ORTHOPAE, P21
[10]
HERBERTS P, 1989, CLIN ORTHOP RELAT R, V249, P48