CLINICAL-EVALUATION OF AN AUTOMATIC SENSITIVITY ADJUSTMENT FEATURE IN A DUAL CHAMBER PACEMAKER

被引:15
作者
WILSON, JH
LOVE, CJ
WETTENSTEIN, EH
机构
[1] OHIO STATE UNIV,COLUMBUS,OH 43210
[2] INTERMED INC,FREEPORT,TX
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1990年 / 13卷 / 10期
关键词
automatic sensitivity adjustment; dual chamber pacemaker;
D O I
10.1111/j.1540-8159.1990.tb02017.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The clinical evaluation of a new automatic sensitivity adjustment feature in the Cosmos fl, Model #284–05 is described. This feature, designed to maintain a 2:1 safety margin for sensing intrinsic signals, was activated by way of special programmer software in ten patients at two centers. While the feature worked satisfactorily in some cases, it did not perform as expected, and undersensing in both chambers was observed. This may have been due to biasing the adjustment toward maximum rather than minimum electrograms, This study suggested that electrograms may vary by more than 100% which underscores the importance of this feature, and reinforces the need for continued development. Copyright © 1990, Wiley Blackwell. All rights reserved