TECHNETIUM 99M RADIOLABELING OF AEROSOLIZED DRUG PARTICLES FROM METERED DOSE INHALERS

被引:7
作者
AUG, C [1 ]
PERRY, RJ [1 ]
SMALDONE, GC [1 ]
机构
[1] SUNY STONY BROOK,DEPT MED,DIV PULM CRIT CARE,STONY BROOK,NY 11794
来源
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG | 1991年 / 4卷 / 02期
关键词
CASCADE IMPACTION; RADIOISOTOPIC; GRAVIMETRIC; AND HPLC PARTICLE SIZE DISTRIBUTIONS; METERED DOSE INHALERS;
D O I
10.1089/jam.1991.4.127
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
To assess mechanisms of bronchodilation and effectiveness of metered dose inhalers, it may be useful to determine sites of drug deposition in the lung. To establish suitable test aerosols, two brands of metered dose inhalers containing bronchodilator (Brethaire, Proventil) were radiolabeled with technetium (Tc-99m) and tested to determine if the distribution of radioisotope in the aerosol was representative of the distribution of agonist activity. Cascade impaction was used to determine the particle size distribution of the radioisotope and drug aerosols by assaying each stage of the cascade using scintillation and HPLC techniques. Possible influences of the radiolabeling method and delivery techniques on the particle distribution were assessed by analyzing distributions from non-radiolabeled inhalers using HPLC. For these drugs, there was an excellent correlation between the distribution of radioactivity and the drug within the captured aerosol (Brethaire r = 0.994, Proventil r = 0.998, 20 - 200 consecutive puffs). Distributions were close to log-normal and differences in mass median aerodynamic diameter (MMAD) between the radioisotope and agonist activities were not significant (Brethaire, MMAD +/- sigma g, radiolabel vs drug = 4.7 +/- 2.1-mu-m, vs 4.4 +/- 1.7-mu-m, and Proventil, 2.5 +/- 2.1-mu-m, vs 2.4 +/- 2.0-mu-m. Non-labeled inhalers produced similar drug distributions (Brethaire, 4.2 +/- 1.8-mu-m, and Proventil 2.0 +/- 1.9-mu-m). Pausing between actuations resulted in slightly smaller distributions (Brethaire, 3.6 +/- 1.8-mu-m, Proventil 1.8 +/- 1.8-mu-m, 20 puffs-60 sec pauses) but the differences were not significant. In addition, to search for multimodal distributions and assess the accuracy of the MMAD measurement via our cascade impactor (Delron), we also measured the distribution of the mass of material within the aerosols using a weight-sensitive cascade (California Measurements). Using the latter device (1 puff), the mass distributions of both aerosols were similar to the values obtained from the puffs with pauses (Brethaire 3.8 +/- 2.3-mu-m, Proventil 1.4 +/- 2.2-mu-m). Multi-modal distributions were not found. By all assessments, the distributions were nearly log-normal with drug activity well described by the radiolabel.
引用
收藏
页码:127 / 138
页数:12
相关论文
共 5 条
[1]  
FULLER H.D., DOLOVICH M.B., POSMITUCK G., WONG PACK W., NEWHOUSE M.T., Pressurized. Aerosol versus Jet Aerosol Delivery to Mechanically Ventilated Patients, Am. Rev. Respir. Dis., 141, pp. 440-444, (1990)
[2]  
KOHLER D., FLEISCHER W., MATTHYS H., New Method for Easy Labeling of Beta-2-Agonists in the Metered Dosage Inhaler with Technetium 99m, Respiration, 53, pp. 65-73, (1988)
[3]  
SHORT M.D., SINGH C.A., FEW J.D., STUDDY P.R., HEAF P.J.D., SPIRO S.G., Labelling and Monitoring of Lung Deposition of an Inhaled Synthetic Anticholinergic Bronchodilating Agent, chest., 80, pp. 918-921, (1981)
[4]  
SMALDONE G.C., PERRY R.J., DEUTSCH D.G., Characteristics of Nebulizers Used in the Treatment of AIDS-Related Pneumocystis Carrnii Pneumonia, J. Aerosol Med., 1, 2, pp. 113-126, (1988)
[5]  
SUMMERS Q.A., CLARK A.R., HOLLINGWORTH A., FLEMING J., HOLGATE S.T., The Preparation of a Radiolabelled Aerosol of Nedocromil Sodium for Admiinstration by Metered-Dose Inhaler that Accurately Preserves Particle Size, Drug Invest., 2, pp. 1-9, (1990)